Wireless Assessment of Respiratory and Circulatory Distress in Chronic Obstructive Pulmonary Disease - Validation Study
Overview
- Phase
- Not Applicable
- Intervention
- Not specified
- Conditions
- Chronic Obstructive Pulmonary Disease
- Sponsor
- University Hospital Bispebjerg and Frederiksberg
- Enrollment
- 20
- Locations
- 1
- Primary Endpoint
- Comparison of pulse rate (beats per minute) measurement with Nonin WristOx 3150 vs standard monitoring (Phillips IntelliVue)
- Status
- Completed
- Last Updated
- 5 years ago
Overview
Brief Summary
For patients admitted to the medical ward, it is often difficult to predict if their clinical condition will deteriorate, however subtle changes in vital signs are usually present 8 to 24 hours before a life-threatening event such as respiratory failure leading to ICU admission, or unanticipated cardiac arrest. Such adverse trends in clinical observations can be missed, misinterpreted or not appreciated as urgent. New continuous and wearable 27/7 clinical vital parameter monitoring systems offer a unique possibility to identify clinical deterioration before patients condition progress beyond the point-of-no-return, where adverse events are inevitable. As part of the WARD-COPD project, this validation study aim to assess the accuracy of physiologic parameters derived from standard and wireless patient monitors
Investigators
Mikkel Elvekjaer
MD, Principal Investigator
University Hospital Bispebjerg and Frederiksberg
Eligibility Criteria
Inclusion Criteria
- •Adult patients admitted with AECOPD
Exclusion Criteria
- •Patient expected not to cooperate
- •Patient allergic to plaster, plastic or silicone
- •Patients with pacemaker or implantable cardioverter-defibrillator (ICD) unit
Outcomes
Primary Outcomes
Comparison of pulse rate (beats per minute) measurement with Nonin WristOx 3150 vs standard monitoring (Phillips IntelliVue)
Time Frame: Two hours of monitoring (measurement interval of 15 mins)
Analysed using a Bland Altman (BA) analysis to measure mean difference and limits of agreement between the two methods. The BA analysis will correct for repeated measures within each subject if required.
Comparison of peripheral oxygen saturation (percent) measurement with Nonin WristOx 3150 vs standard monitoring (Phillips IntelliVue)
Time Frame: Two hours of monitoring (measurement interval of 15 mins)
Analysed using a Bland Altman (BA) analysis to measure mean difference and limits of agreement between the two methods. The BA analysis will correct for repeated measures within each subject if required.
Comparison of heart rate (beats per minute) measurement with Isansys Lifetouch patch vs standard monitoring (Phillips IntelliVue)
Time Frame: Two hours of monitoring (measurement interval of 15 mins)
Analysed using a Bland Altman (BA) analysis to measure mean difference and limits of agreement between the two methods. The BA analysis will correct for repeated measures within each subject if required.
Comparison of respiratory rate (breaths per minute) measurement with Isansys Lifetouch patch vs standard monitoring (Phillips IntelliVue) vs direct observation
Time Frame: Two hours of monitoring (measurement interval of 15 mins)
Analysed using a Bland Altman (BA) analysis to measure mean difference and limits of agreement between the three methods. The BA analysis will correct for repeated measures within each subject if required.
Comparison of systolic and diastolic blood pressure (mmHg) measurement with Meditech BP-05 vs standard monitoring (Phillips IntelliVue)
Time Frame: Two hours of monitoring (measurement interval of 15 mins)
Analysed using a Bland Altman (BA) analysis to measure mean difference and limits of agreement between the two methods. The BA analysis will correct for repeated measures within each subject if required.