临床试验/NCT06829615
NCT06829615已完成不适用

Contactless Measurement of Heart Rate, Heart Rate Variability, Breathing Rate and Blood Pressure Using Face Video Analysis (Remote Photoplethysmography) - Validation of Shen.AI Software

SHEN.AI spółka z ograniczoną odpowiedzialnością oddział w Polsce4 个研究点 分布在 1 个国家目标入组 132 人开始时间: 2025年5月6日最近更新:

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
132
试验地点
4
主要终点
Accuracy of the average heart rate (HR) measurement using the device under test compared to the reference measurement.

研究概览

简要总结

The purpose of this study it to evaluate the accuracy of heart rate, heart rate variability, breathing rate and blood pressure measurement with the use of Shen.AI software.

详细描述

The primary objective of this clinical study is to assess the accuracy of heart rate (HR), heart rate variability (HRV), breathing rate (BR), and blood pressure (BP) measurements obtained using Shen.AI software as the investigational device under its intended conditions of use, enabling users to obtain reliable information about their physiological state. These conditions include measurements taken at rest, in a home-like environment, with participants independently operating a mobile device or laptop equipped with a camera.

Physiological parameters, including HR, HRV, BR, and BP, will be measured and estimated using remote photoplethysmography (rPPG) based on a 1-minute facial video analysis processed by Shen.AI software.

Throughout all measurements, participants will be connected to a reference device that records ECG while simultaneously measuring chest impedance (impedance pneumography). Additionally, each participant will wear a respiratory belt transducer. Depending on the measurement type, the participant's blood pressure will be assessed using either an auscultatory or an automatic blood pressure monitor. The study will be conducted on a diverse group of participants, varying in blood pressure, gender, age, skin phototype, and BMI.

For each measured parameter, results obtained from both methods will be compared using Bland-Altman analysis or non-parametric limits of agreement.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Diagnostic
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Age ≥ 18 years
  • Ability to read and write in Polish or English
  • Ability to operate a mobile device and a laptop without requiring glasses
  • Ability to understand the study procedure
  • Ability to communicate and follow instructions from the study personnel
  • Provision of written informed consent for participation in the study and for the processing of personal data by the research team and the study sponsor
  • Blood pressure values in accordance with the distribution of the study group

排除标准

  • Pregnancy
  • Age below 18 years
  • Failure to provide the required information in the participant's informational questionnaire
  • Non-compliance with the study procedure or instructions from the study personnel
  • Extensive facial skin damage (e.g., abrasions, wounds, burns)
  • A disease process affecting a significant portion of the face
  • Extensive facial dressing
  • Extensive facial tattoos or facial painting
  • Significant facial deformity (e.g., tumor)
  • Inability to maintain a stable head position during measurement
  • Persistent cough preventing stillness during measurement
  • Respiratory dysfunction such as dyspnea, irregular or shallow breathing
  • Diagnosed arrhythmia (except for sinus bradycardia or sinus tachycardia)
  • Moderate or severe anemia
  • Diagnosed conditions such as left ventricular systolic dysfunction, aortic valve stenosis, or other cardiac or circulatory diseases leading to low stroke volume, low pulse amplitude, or pulsus paradoxus
  • Implanted cardiac pacemaker
  • Blood pressure difference between arms exceeding 15 mmHg for systolic pressure or 10 mmHg for diastolic pressure
  • Inability to determine blood pressure in both arms

结局指标

主要结局

Accuracy of the average heart rate (HR) measurement using the device under test compared to the reference measurement.

时间窗: One appointment (day 1)

Average HR will be estimated using Shen.AI software and compared with simultaneously recorded ECG .

Accuracy of the measurement of the heart rate variability index - Standard Deviation of Normal-to-Normal intervals (SDNN) using the device under test compared to the reference measurement.

时间窗: One appointment (day 1)

The heart rate variability index (SDNN) will be measured using Shen.AI software and compared with the simultaneously recorded ECG.

Accuracy of the breathing rate (BR) measurement using the device under test compared to the reference measurement.

时间窗: One appointment (day 1)

Average BR will be estimated using Shen.AI software and compared with the recorded impedance pneumography signal in the absence of read-out capability supported by the respiratory belt transducer.

Accuracy of systolic and diastolic blood pressure (BP) measurements using the device under test in relation to the reference measurement.

时间窗: One appointment (day 1)

Average BP will be estimated using Shen.AI software and compared with auscultatory blood pressure measurements.

Accuracy of systolic and diastolic blood pressure (BP) measurement using the device under test, taking into account the calibration of the BP measurement function, in relation to the reference measurement.

时间窗: Two appointments (14 ± 3 days apart)

Average BP will be estimated using Shen.AI software and compared with auscultatory blood pressure measurements. Measurements will be performed during two visits, with the second visit occurring 14 ± 3 days apart.

次要结局

  • Accuracy of the measurement of other heart rate variability metrics - Root Mean Square of Successive Differences (RMSSD) using the device under test.(One appointment (day 1))
  • Accuracy of blood oxygen saturation measurement using the device under test(One appointment (day 1))
  • Accuracy of Body Mass Index (BMI) estimation using the device under test.(One appointment (day 1))

研究者

发起方
SHEN.AI spółka z ograniczoną odpowiedzialnością oddział w Polsce
申办方类型
Industry
责任方
Sponsor