A Comparative Study Of Three Blood Pressure Measuring Techniques In Neurosurgical Patients Undergoing Intraoperative Blood Pressure Manipulation
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 40
- 试验地点
- 1
- 主要终点
- Non invasive blood pressure monitoring is not reliable alternative to invasive arterial blood pressure monitoring during intraoperative blood pressure manipulation.
研究概览
简要总结
This is prospective observational study in patients undergoing neurosurgical procedure requiring intraoperative blood pressure manipulation between 2014 and 2016 in neurosurgery operation theatre of LTGM Hospital and Medical college, Sion, Mumbai. During study we compared readings of three blood pressure measuring technique namely Invasive direct arterial blood pressure , non invasive oscillometric blood pressure and auscultatory manual blood pressure at different ranges of blood pressures during intraoperative blood pressure manipulation and establish correlation between these techniques with reliability and accuracy of non invasive blood pressure measuring techniques (oscillometric and auscultatory) considering invasive arterial line as gold standard for blood pressure measurement.
We find that both the. Non invasive blood pressure measurement techniques tend to systolic blood pressure measurement, the difference was -9 to +21 mm Hg for oscillometric technique and -10 to + 20 mm Hg by auscultatory technique.
Systolic blood pressure at range of 100 to 140 mm Hg both the non invasive techniques are found to be reliable. While at blood pressure above 140 mm Hg both non invasive techniques tend to underestimate systolic pressure reading in which oscillometric by 9 +/- 1.9 mm Hg and auscultatory by 8.58 +/- 0.36 mm Hg which is found to be statistically significant, when we further analyses blood pressure readings below 90 mm Hg. Oscillometric but not auscultatory tended to overestimated systolic pressure reading by 5.7 +/- 3.17 mm Hg.
On analysing diastolic blood pressure both the non-invasive techniques tended to overestimate diastolic pressure, oscillometric by 6.42 +/- 7.60 mm Hg and auscultatory by 6.54 +/- 7.52 mm Hg.
Mean blood pressure readings were found to be reliable with non-invasive technique.
研究设计
- 研究类型
- Observational
入排标准
- 年龄范围
- 18.00 Year(s) 至 65.00 Year(s)(—)
- 性别
- All
入选标准
- •Patients Undergoing Neurosurgical Procedures requiring blood pressure manipulations under anaesthesia.
- •patients subjected to radial artery cannulation.
排除标准
- •1.Severely hemodynamic compromised patients.
- •2.Patients with Arrhythmias.
- •3.Patents in which there is inability to take Radial Artery Cannulation.
- •4.Any Contraindications to Cuff application or inflation.
- •5.Pregnant Females.
- •6.Difference of more than 10 mm of hg blood pressure in both the arms 7.Different surgical positions other than supine.
结局指标
主要结局
Non invasive blood pressure monitoring is not reliable alternative to invasive arterial blood pressure monitoring during intraoperative blood pressure manipulation.
时间窗: Outcome will be assessed after collection of data from all the 40 patients within period of 18 months.
次要结局
- Non invasive blood pressure measuring technique tend to underestimate hypotension and overestimate hypertension(Outcome will be assessed after collection of data from all the 40 samples within 18 month period.)