MedPath

Clinical performance evaluation of the landmark photo checker in upper gastrointestinal endoscopy

Not Applicable
Conditions
on-disabled individuals
Registration Number
JPRN-UMIN000052643
Lead Sponsor
Foundation for Detection of Early Gastric Carcinoma
Brief Summary

The sensitivity of the landmark was 85.7%. The specificity and accuracy rate, were 72.5% and 79.0%, respectively. The individual sensitivities (%) and specificities (%) for the seven landmarks were as follows: esophagogastric junction, 88.9 and 86.7; greater curvature of the cardia, 71.9 and 32.3; lesser curvature of the cardia, 84.4 and 53.1; greater curvature of the body, 90.6 and 90.3; lesser curvature of the body, 90.0 and 71.0; angle of the stomach, 86.2 and 87.5; and pylorus, 89.3 and 87.1.

Detailed Description

Not available

Recruitment & Eligibility

Status
Complete: follow-up continuing
Sex
All
Target Recruitment
64
Inclusion Criteria

Not provided

Exclusion Criteria

Patients after gastric surgery.

Study & Design

Study Type
Interventional
Study Design
Not specified
Primary Outcome Measures
NameTimeMethod
Secondary Outcome Measures
NameTimeMethod
© Copyright 2025. All Rights Reserved by MedPath