Pilot Study for the Validation of the BC3 Bioelectrical Impedance Hydration Factor
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 360
- 试验地点
- 1
- 主要终点
- Estimated Percent Body Fat for Men Aged 18-80
研究概览
简要总结
This was a pilot study to validate the BC3 hydration factor which will be important for determining how well this value represents hydration status compared to assessment of total body water by D20.
详细描述
The BC3 body composition analyzer is manufactured by Stayhealthy. It is a handheld, noninvasive bioelectrical impedance device. When held, the BC3 sends an electrical impulse through the body to determine body fat, lean mass, and total body hydration. Bio-impedance devices measure the change in electrical resistance in body tissues by using a detectable electrical signal that passes through the body. The method is based on the principle that lean body mass contains virtually all the water and conducting electrolytes in the body and provides a good electrical pathway. In contrast, fat or fat-containing tissues produce a poor electrical pathway. By inducing a low-energy, high-frequency signal, a measurement of the baseline resistance to flow of electrical current can be assessed. The resistance measurement relates directly to the volume of the conductor. Through repeated measurements and identifying other parameters (i.e. gender, weight, age and height) proprietary equations were developed that determine total body water, lean mass, and % body fat.
The purpose of this study is to validate the calculated hydration factor produced by the BC3 bioelectrical impedance device. If this value is valid and comparable to the gold standard (D20) then this hydration factor will offer a very quick and inexpensive method to determine total body hydration status using a hand-held bioelectrical impedance device.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Health Services Research
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 80 Years(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Men and women 18-80 years of age
排除标准
- •Acute illness (infection, recent surgery, trauma).
- •Pregnancy
- •Pacemaker or defibrillator
研究组 & 干预措施
deuterium oxide and BC3
Testing of total body water on the BC3 compared to total body water measured by deuterium oxide.
干预措施: D2O (Device)
结局指标
主要结局
Estimated Percent Body Fat for Men Aged 18-80
时间窗: One time point (only one day visit)
The body fat percentage as estimated from BC3 and as calculated by a full body DXA.
Percent Body Fat in Women Aged 18-80 Years Old
时间窗: One time point (only one day visit)
Body fat percentage was measured via Stayhealthy BC3 as well as a full body DXA
Estimated Body Fat % for Children Aged 10-17 Years Old
时间窗: One time point (only one day visit)
Stayhealthy BC3 was analyzed against hydrostatic weighing for children aged 10-17 years old