The Evaluation of a Respiratory Monitor in Surgical Patients With a BMI >35 Undergoing Elective Surgery Under General Anesthesia
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 90
- 试验地点
- 1
- 主要终点
- ExSpiron monitor accurately reflects TV, MV and RR in obese patients in the preoperative holding area before surgery
研究概览
简要总结
This study proposes to evaluate the clinical applicability of the ExSpiron Respiratory Volume Monitor (RVM, ExSpiron™, Respiratory Motion, Inc.; Waltham, MA) in obese surgical patients undergoing general anesthesia. Previous work has demonstrated the ability of the ExSpiron monitor to provide non-invasive, real-time, continuous measurements of respiratory parameters such as tidal volume (TV), minute ventilation (MV) and respiratory rate (RR) mostly in normal weight patients but those studies did not specifically look at obese subjects. Respiratory depression, in the postoperative setting due to residual anesthetics and/or opioid administration, continues to be a significant cause of adverse outcomes. Obese patients are at increased risk for respiratory complications. Currently, there is no objective measure of early respiratory indicators for developing respiratory compromise. Current respiratory assessment in non-intubated patients relies on oximetry data and subjective clinical assessment. Pulse oximetry has been extremely helpful in recognizing oxygen desaturations but it is a late indicator of respiratory decline. There is no current device capable of giving real time ventilatory information such as tidal volume and minute ventilation of a patient that is not mechanically ventilated. The ExSpiron system utilizes an impedance based technology and proprietary algorithms (Respiratory Motion Inc.) to obtain these measurements. The study hypotheses are that the non-invasive, impedance-based RVM monitor will accurately reflect TV, RR and MV in obese surgical patients before induction of general anesthesia, during controlled ventilation and following extubation; that ExSpiron will accurately reflect the post-extubation respiratory status of the patient; and that apnea and hypopnea episodes in the recovery room as detected by the ExSpiron monitor are correlated with the individual risk for obstructive sleep apnea as determined by the STOP-Bang risk stratification.
研究设计
- 研究类型
- Observational
- 观察模型
- Cohort
- 时间视角
- Prospective
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •ASA 1 - 3 patients with a body mass index (BMI) ≥ 35kg/m2 presenting for elective surgery under general anesthesia aged > 18 years. Ability to provide written informed consent.
排除标准
- •Emergent surgical patients, patients with a BMI < 35kg/m2, patients not competent to give informed consent, pregnant patients (women of child bearing potential who want to participate will have a pregnancy test done prior to enrollment in the study), patients with implantable electronic devices (pacemakers, stimulators, etc.); predictable interference with surgical procedure or standard of care.
结局指标
主要结局
ExSpiron monitor accurately reflects TV, MV and RR in obese patients in the preoperative holding area before surgery
时间窗: 1 - 1.5 hours before surgery
Patient will have a monitoring strip applied to right chest in the pre op holding area, and respiratory data will be recorded to the monitor. These will be compared with data from a manual handheld spirometer via 5 recordings over a time period of 5 minutes.
次要结局
- ExSpiron will accurately reflect the post-extubation respiratory status of the patient(1 - 3 hours in the recovery room)