Brief Summary

The "ball-in-socket" design of the Medial-Pivot knee system (MicroPort Orthopedics, Arling-ton, Tennessee, USA) aims to reproduce normal knee kinematics by medializing its rotational axis. Thegoal of this study was to measure knee range of motion (ROM) with this implant after a mean follow-up of 10 years and to report the survivorship and long-term clinical and radiological outcomes. Wehypothesized the prosthetic knee would have at least 120◦flexion at 10 years. This was retrospective, single-centre study of 74 Medial-Pivot knees implanted in71 patients (average age of 69 years) between May 2005 and November 2007. All patients who receiveda Medial-Pivot knee were included consecutively. The mean follow-up was 10 years. Clinical and radio-logical assessments were performed using the Knee Society Score (KSS) and Ewald's score. Kaplan-Meirsurvival analysis was used to calculate survivorship.

Detailed Description

methodsThis was retrospective, single-centre study of 74 Medial-Pivotknees implanted in 71 patients between May 2005 and Novem-ber 2007. During this period, several types of semi-constrainedprimary TKA implants were used.The "ball-in-socket" design of the Medial-Pivot knee system (MicroPort Orthopedics, Arling-ton, Tennessee, USA) aims to reproduce normal knee kinematics by medializing its rotational axis. Thegoal of this study was to measure knee range of motion (ROM) with this implant after a mean follow-up of 10 years and to report the survivorship and long-term clinical and radiological outcomes. Wehypothesized the prosthetic knee would have at least 120◦flexion at 10 years. This was retrospective, single-centre study of 74 Medial-Pivot knees implanted in71 patients (average age of 69 years) between May 2005 and November 2007. All patients who receiveda Medial-Pivot knee were included consecutively. The mean follow-up was 10 years. Clinical and radio-logical assessments were performed using the Knee Society Score (KSS) and Ewald's score. Kaplan-Meirsurvival analysis was used to calculate survivorship.

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
71
Inclusion Criteria
  • patients > 18 years
  • patienst must be operated
  • patient with mid-line skin incision
Exclusion Criteria
  • patients < 18 years

Study & Design

Study Type
OBSERVATIONAL
Study Design
Not specified
Primary Outcome Measures
NameTimeMethod
Flexion with an Advance® Medial-Pivot knee2years

The objective of this study was to evaluate the flexion in patients with an Advance® Medial-Pivot knee after an average follow-upof 10 years. Knee flexion was measured with a goniometer in the office; thegreater trochanter, knee centre and lateral malleolus were used asreference points.

Secondary Outcome Measures
NameTimeMethod