Desaturation Validation of INVSENSOR00039

Completed
Conditions
Healthy
Registration Number
NCT04112784
Lead Sponsor
Masimo Corporation
Brief Summary

This study is designed to compare the accuracy of a noninvasive measurement of oxygen saturation compared to reference values obtained by a laboratory blood gas analyzer. Arterial blood samples will be collected from healthy adult subjects while undergoing a desaturation procedure wherein the concentration of oxygen inhaled is slowly reduced. The subject will then be returned to breathing room air.

Detailed Description

Not available

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
38
Inclusion Criteria

Not provided

Exclusion Criteria

Not provided

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Primary Outcome Measures
NameTimeMethod
Oxygen Saturation (SpO2) Accuracy of Sensor1-5 hours

Performance of the sensors will be determined by comparing the noninvasive oxygen saturation measurement (SpO2) of the pulse oximeter sensors to the arterial oxygen saturation (SaO2) value obtained from a reference blood sample and calculating the Accuracy root square mean (ARMS) value.