MedPath

Effect of a Pre Heparin Coated Dialysis Filter on Coagulation During Hemodialysis

Phase 4
Completed
Conditions
Renal Failure
Interventions
Device: a pre-heparin-coated hemodialysis filter; Hepran
Registration Number
NCT01388270
Lead Sponsor
Oslo University Hospital
Brief Summary

The purpose of this study is to determine whether the heparin-coated dialysis filter is superior to a conventional filter during hemodialysis in twelve stable, chronic hemodialysis (HD) patients.

The antithrombogenic properties of the heparin-coated dialysis filter and of the conventional filter will be compared by means of statistical tests.

Detailed Description

In the present study twelve chronic hemodialysis (HD) patients will be included. Exclusion criteria are use of acetylsalicylic acid and/warfarin, clinical signs of infection and disseminated malignant disease. For each patient four HD sessions will be investigated; alternating with use of the pre-heparin-coated filter or the conventional filter. Dalteparin is reduced to 50% of the conventional dose and is given as one single bolus dose at start of HD.

The venous drip chamber will be visually inspected for evaluation of clot score during each HD session. The filter will be visually inspected for clot score at the end of each session.

Blood specimens will be drawn systematically during each HD session to evaluate platelet and coagulation activity as well as anti Factor Xa activity, haematology and urea/creatinine to evaluate dialysis effect.

The antithrombogenic properties of the pre-heparin-coated filter and the conventional filter will be compared by means of statistical tests. The hypothesis is that the pre-heparin-coated filter induces significantly less clot score in the extracorporeal system during hemodialysis and also significantly less intravascular activation of the coagulation system compared to the conventional filter.

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
12
Inclusion Criteria
  • Stable HD patients of at least 18 years of age.
  • HD treatment for at least one month.
  • Dialysis time at least 4 hours three times per week.
  • Stable Hemoglobin value of at least 11 g/dl during the last 4 weeks.
Exclusion Criteria
  • Use of warfarin and/or acetylsalicylic acid.
  • Disseminated malignant disease.
  • Clinical signs of infection.

Study & Design

Study Type
INTERVENTIONAL
Study Design
CROSSOVER
Arm && Interventions
GroupInterventionDescription
Evodiala pre-heparin-coated hemodialysis filter; Heprana pre-heparin-coated hemodialysis filter
Primary Outcome Measures
NameTimeMethod
Activation of the intravascular coagulation system during hemodialysis with use of a pre-heparin-coated filter compared to a conventional filterActivation of the intravascular coagulation system is assessed during four dialysis sessions within a two weeks period for a single patient

Blood specimens for analysis of prothrombin fragment 1+2 and beta-thromboglobulin will be drawn at start of hemodialysis and after three and four hours of all hemodialysis sessions.

Secondary Outcome Measures
NameTimeMethod
A secondary outcome measure is degree of clot score in the extracorporeal system during hemodialysis with use of a pre-heparin-coated filter compared to the conventional filter.The degree of clot score in the extracorporeal system is assessed during four hemodialysis sessions that for a single patient take place within a time frame of two weeks

Degree of clot score will be evaluated in the venous drip chamber as follows: 1=normal (no signs of clot). 2=fibrinous ring, 3=clot and 4=stop in dialysis due to high degree of clotting

Trial Locations

Locations (1)

Department of nephrology, Oslo University Hospital, Ullevål

🇳🇴

Oslo, Norway

© Copyright 2025. All Rights Reserved by MedPath