Remote Blood Pressure Monitoring With the Wearable SENBIOSYS Photoplethysmographic Device

Not Applicable
Completed
Conditions
Blood Pressure
Interventions
Device: Senbiosys
Registration Number
NCT04379986
Lead Sponsor
Stéphane Cook, Prof
Brief Summary

During the last couple of years, a growing number of wearable devices evolved to provide accurate, cheap and non-invasive monitoring of vitals parameters.This connected care technology could be helpful for treatment and care during a pandemic such as COVID-19. The use of these non-invasive remote monitoring devices can help health care providers to assess patient's vital signs and symptom progression, reducing reducing patient and healthcare provider contact and exposure to COVID-19 during this pandemic.

Detailed Description

Several cuff-less blood pressure monitoring devices have emerged in patient care using photoplethysmography. PPG utilizes an infrared or visible light to measure peripheral volumetric variations of blood circulation and its waveform has been proven to have a good correlation with BP waveform.Nonetheless, the accuracy of these instruments is of fundamental importance.

Several validation procedures for assessing the precision of BP monitoring devices have been developed, including protocols by the Association for the Advancement of Medical Instrumentation, International Organization for Standardization and European Society of Hypertension. Adherence to these guidelines are essential to ensure the effectiveness of BP measurement of the wearable devices.

Therefore, the investigators propose to test wearable low-noise, low-power SENBIOSYS photoplethysmography signals for estimation of BP.

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
35
Inclusion Criteria

Not provided

Exclusion Criteria

Not provided

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Arm && Interventions
GroupInterventionDescription
Standardized measures in the intensive care unitSenbiosysInvasive BP monitoring is a commonly used technique in the ICU and is used to guide many intensive care unit therapies. Enrolled patients must have had an arterial catheter in place at the time of inclusion to the study. Arterial catheterization will be performed by the intensive care team according to current medical guidelines. No arterial catheters were placed for the sole purpose of this study. The Senbiosys device will be placed on the opposite arm of the arterial catheter for simultaneous measurements.
Standardized measures in the cardiac catheterization laboratorSenbiosysImmediately after coronary angiography, intra-arterial BP waveforms will be recorded using a fluid-filled catheter via right femoral or radial access. The catheter will be flushed before any waveform recordings is made. At first, the catheter will be positioned in the aorta for 3 minutes of stable BP waveforms recording. Intracoronary nitroglycerin will be administered at a dose of 300 µg, newly preceding 3 minute of recording. At the end of the coronary angiography, additional 3 minutes of recording will be performed in the aorta.
Primary Outcome Measures
NameTimeMethod
Feasibility of non-invasive BP monitors: blood pressure10 to 15 minutes per patient (once)

The primary outcome is the assessment of mean bias (± 95% CI or precision of bias) for systolic blood pressure (SBP), diastolic blood pressure (DBP) and mean blood pressure (MBP) between invasive and non-invasive BP measurements. The standard deviation of the bias (±95% limits of agreement), will be assessed for SBP, DBP, and MBP measurements.

Secondary Outcome Measures
NameTimeMethod
Percentage of signal with artefact,10 to 15 minutes per patient (once)

Secondary outcomes include reliability index (Qualification Index QI) for blood pressure epochs and count of qualified epochs.

Trial Locations

Locations (1)

University and hospital Fribourg

🇨🇭

Fribourg, Switzerland