Clinical Trials/NCT04324385
NCT04324385CompletedNot Applicable

Accuracy of Pulse Oximeters With Profound Hypoxia

Compumedics Limited1 site in 1 country12 target enrollmentStarted: April 9, 2019Last updated:
Conditions

Trial Snapshot

Phase
Not Applicable
Status
Completed
Enrollment
12
Locations
1
Primary Endpoint
Testing the accuracy of a new pulse oximeter device

Study Overview

Brief Summary

To test the accuracy of a non-invasive blood oxygen saturation monitor which is placed on a subject with profound hypoxia.

Detailed Description

A radial arterial cannula was placed in either the left or right wrist of each subject for arterial blood sampling and blood pressure monitoring. Blood gas analysis to determine oxyhemoglobin saturation (SaO2)was performed using an ABL-90 multi-wavelength oximeter (Hemoximeter, Radiometer, Copenhagen, serial 1393-090R0359N0002). This instrument contains factory certified calibration standards and quality control algorithms.

Each subject had two control blood samples taken at the beginning of each experiment, while breathing room air. Hypoxia was then induced to different and stable levels of oxyhemoglobin saturation (between 70-100%) by having subjects breathe mixtures of nitrogen, room air, and carbon dioxide. Each plateau level of oxyhemoglobin saturation was maintained for at least 30 seconds or until reference pulse oximeters readings were stable. Two arterial blood samples were then obtained, approximately 30 seconds apart. Each stable plateau therefore was maintained for at least 60 seconds with SpO2 fluctuating by less than 2-3%. The plateaus were nominally at 100%, room air saturation, 93%, 90%, 87%, 85%, 82%, 80%, 77%, 75% and 70%. A total of 270 samples were obtained at the saturation plateaus across this span. Data were provided for analysis. At least 200 data points were collected for each type of oximeter and probe combination studied.

Pulse oximeter data is taken as 5 second averages corresponding to the point of arterial blood analysis. Individual data points may be missed or excluded for dropped signals or failure of the oximeter signal to achieve an appropriate plateau. Data is plotted as Hemoximeter data (SaO2) vs. pulse oximeter bias (SpO2 - SaO2). A different marker is used for each study subject. Linear regression is shown for all subjects combined, and the equation with R2 is shown on the plot. Mean bias is displayed as a solid horizontal line, and the upper and lower limits of agreement (mean bias ± 1.96•SD*) are shown by dashed horizontal lines. For the "pooled" plots, different markers are used for each pulse oximeter. Tables of mean, standard deviation, standard error, minimum, maximum, 95% confidence interval, count and root mean square are provided for each oximeter's bias, and all oximeters combined in the following ranges of SaO2 (Hemoximeter): 60 - 80%, 80 - 100%, 60 - 100%, 70 - 100%, 50 - 60%, 60 - 70%, 70 - 80%, 80 -90%, and 90 - 100%.

Study Design

Study Type
Observational
Observational Model
Cohort
Time Perspective
Prospective

Eligibility Criteria

Ages
21 Years to 49 Years (Adult)
Sex
All
Accepts Healthy Volunteers
Yes

Inclusion Criteria

  • Normal hemoglobin levels (Hemoglobin ≤ 10gm•dl-1)
  • Non-smoking individuals

Exclusion Criteria

  • Unhealthy

Outcomes

Primary Outcomes

Testing the accuracy of a new pulse oximeter device

Time Frame: Approx. 60 mins per subject

The primary outcome measure was to find the agreement between methods of measurement with multiple observations per individual. Spo2 readings from the test device were compared with a factory certified and calibrated oxyhemoglobin saturation monitor (i.e. ABL-90 multi-wavelength oxiemter) and two other standard oximeter devices using root mean squared error as the measure of accuracy.

Secondary Outcomes

No secondary outcomes reported