A Study of Agreement Between Oxygen Saturation Measured by Two Probes in Infants on the Neonatal Unit

Not Applicable
Completed
Conditions
Neonatal Disease
Interventions
Device: Oxygen saturation probe Mindray 520 N and Mindray 521 N
Registration Number
NCT05065229
Lead Sponsor
University of Nottingham
Brief Summary

On the neonatal unit in Derby, two types of probes are in routine use: Mindray 520 N and Mindray 521 N. Feedback from parents and staff show that there are concerns that the two probes given very different oxygen saturation readings and, often, do not agree with each other. This has raised concerns that infants' clinical care may be affected by the choice of probe. It is important to know if the two probes give similar results or not to ensure that infants get the appropriate monitoring and respiratory support as needed for good neonatal care. In this study, we will compare the reading made by the two probes and determine whether the readings of the two probes agree with each other and if any disagreement is such that clinical decision making is affected by the difference.

Detailed Description

Continuous monitoring of oxygen saturation level is a key component of neonatal intensive care and NICE recommends that all neonatal services "ensure that systems are in place for preterm babies to have a target saturation level of 91% to 95%" and that this is monitored using "continuous pulse oximetry" (NICE Quality statement 4 in Quality Standard QS193: https://www.nice.org.uk/guidance/qs193/chapter/Quality-statement-4-Oxygen-saturation). Oxygen saturation monitoring is used to provide optimal respiratory support to preterm infants to ensure adequate oxygenation of tissues while preventing risks of conditions such as retinopathy of prematurity and bronchopulmonary dysplasia \[1\]. It is equally important in term born infants who require intensive care such as for determining adequate oxygenation for respiratory support and screening for congenital heart disease \[2\]. For these reasons, most infants in neonatal units are continuously monitored using pulse oximetry, a simple and painless test that measures blood oxygen levels by passing a small beam of light through the infant's hand or foot. This measures the changes in light absorption to calculate the levels of oxygenated and deoxygenated blood. To do this the light (or probe) has to be attached to the infant's hand or foot using a tie or a sticker. The probe must be attached firmly for the saturation monitor to pick up the correct readings. As neonatal skin is delicate and it is often difficult to stick objects to it, neonatal oxygen saturation probes are carefully designed. Many different types of saturation probes are in routine use in neonatal units. It is important to ensure that they give accurate and comparable readings. In the neonatal unit in Derby, two types of probes are in routine use: Mindray 520 N and Mindray 521 N. Feedback from parents and staff show that there are concerns that the two probes given very different oxygen saturation readings and, often, do not agree with each other. This has raised concerns that infants' clinical care may be affected by the choice of probe. It is important to know if the two probes give similar results or not to ensure that infants get the appropriate monitoring and respiratory support as needed for good neonatal care. In this study, we will compare the reading made by the two probes and determine whether the readings of the two probes agree with each other and if any disagreement is such that clinical decision making is affected by the difference.

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
20
Inclusion Criteria
  • Infants who need oxygen saturation monitoring for clinical care
  • Infants born preterm (<37 weeks' gestation) who are < 7 days of age and need respiratory support including supplemental oxygen
  • Infants born preterm (<37 weeks' gestation) who are in supplemental oxygen at >34 weeks' gestation
  • Infants born at term gestation (< 37 weeks' gestation) who require respiratory support including supplemental oxygen
  • Infants with suspected cardiac disease
Exclusion Criteria
  • Infants who have intravenous or arterial lines attached to either foot.
  • Infants who have any condition that may cause differential oxygen saturation or blood flow to the lower limbs such as venous thrombosis.
  • Infants who have local infections or skin lesions on either foot.
  • Infants older than 5 months of age

Study & Design

Study Type
INTERVENTIONAL
Study Design
PARALLEL
Arm && Interventions
GroupInterventionDescription
AOxygen saturation probe Mindray 520 N and Mindray 521 NRight foot no.520 Left foot no.521
BOxygen saturation probe Mindray 520 N and Mindray 521 NRight foot no.521 Left foot no.520
Primary Outcome Measures
NameTimeMethod
oxygen saturation of babies using Minday 520 N and Minday 521 N12 months

Oxygen saturation readings using Minday 520 N and Minday 521 N probes simultaneously

Secondary Outcome Measures
NameTimeMethod
clinical decision making12 months

clinical decisions made, such as giving supplemental oxygen to infants on the neonatal unit following oxygen saturation readings from Minday 520 N and Mindray 521 N

Trial Locations

Locations (1)

University Hospitals of Derby and Burton

🇬🇧

Derby, Derbyshire, United Kingdom