A Concurrence Study of Diagnostic Visualization Comparing Blue Light Imaging (BLI) With Flexible Spectral-Imaging Color Enhancement (FICE)

Fujifilm Medical Systems USA, Inc.2 个研究点 分布在 1 个国家目标入组 20 人开始时间: 2016年8月18日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
20
试验地点
2
主要终点
Image Quality Scores Using Likert Scale

研究概览

简要总结

The study is intended to demonstrate substantial equivalence of visualization using BLI to the predicate device, FICE, so as to support the addition of BLI as an adjunct to white light imaging in the company's video processor.

详细描述

This clinical study will consist of acquisition of images, followed by a reader study.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Diagnostic
盲法
None

入排标准

年龄范围
21 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Referred for clinically indicated gastrointestinal diagnostic endoscopy
  • Ability to understand the requirements of the study, to provide written informed consent and to comply with the study protocol
  • Ability to understand and provide written authorization for the use and disclosure of protected health information (PHI) per the Health Insurance Portability and Accountability Act (HIPAA) privacy ruling.

排除标准

  • Under the age of 21 years
  • Is an inmate (see US Code of Federal Regulations 45 C.F.R. 46.306)

结局指标

主要结局

Image Quality Scores Using Likert Scale

时间窗: Two to three months

Likert score ratings of image quality for Blue Light Imaging (BLI), Blue Light Imaging-bright (BLI-brt) and Flexible spectral-Imaging Color Enhancement (FICE) images. Each BLI and BLI-Brt setting was compared to FICE, to determine which settings were equivalent to FICE. Likert Score Scale: 1. Not seen; 2. poor but usable, characteristic features are detectable but details are not fully reproduced; features just visible; 3. good; allows an adequate assessment, details of anatomical structures are visible but not necessarily clearly defined; details emerging; 4. very good; allows an excellent assessment, anatomical details are clearly defined; details clear This was a blinded and randomized Reader assessment of all BLI and FICE images.

次要结局

未报告次要终点