A Feasibility Study Evaluating the Use of SafeBreak Vascular in the Geriatric Trauma Population
- Conditions
- IV Dislodgement
- Interventions
- Device: SafeBreak Vascular
- Registration Number
- NCT03505957
- Lead Sponsor
- Lineus Medical
- Brief Summary
SafeBreak Vascular is a medical device indicated for fluid flow control during the administration of IV fluids/medication to the patient's vascular system and to aid in the prevention of unintended disruption and dislodgment of IVs. SafeBreak Vascular may be used for any hospitalized patient with gravity tubing or IV pumps, for intermittent infusion or continuous infusion.The primary objective of this study is to determine if the delivery of IV medications and/or fluids with SafeBreak Vascular functions with the same reliability as current standard IV tubing.
- Detailed Description
Study participants will have a SafeBreak Vascular installed in each of their existing or newly installed IV lines. The clinical literature shows in prospective observational studies that a mean of 9.2% of patients dislodge their peripheral IVs. SafeBreak Vascular is a medical device designed to aid in the prevention of unintended disruption and dislodgement of IVs, such as peripheral IVs, peripherally inserted central lines, central lines, etc. SafeBreak IV will be installed in each IV line between the study participant's catheter and the IV tubing that goes to the IV bag or IV pump. When a harmful force is placed on the IV line, SafeBreak Vascular is designed to separate so that the harmful force is removed from the line and IV dislodgement is prevented. SafeBreak Vascular has a valve in each end of the device. Upon separation, each valves closes, stopping the flow of fluid. The primary objective of the study is to determine if the delivery of IV medications and/or fluids with SafeBreak Vascular functions with the same reliability as current standard IV tubing. The number of IV dislodgements and IV restarts will be recorded. The study will also collect data for analysis concerning the following:
* device design characteristics
* certain human factors (patient or operator) associated with the use of the device
* the impact of this device usage on clinician workflow
* any unknown safety concerns
* obtain preliminary data for use in designing a subsequent pivotal study of the device.
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 40
- Geriatric trauma patients (> or equal to age 55)
- Participants able to provide informed consent or have a legally authorized representative (LAR) provide consent
- The patients must have IV access (existing central venous catheter, peripherally inserted central catheter, or peripheral IV's) or need IV access
- The patients must be at least 55 years of age.
- Unable to obtain informed consent or without an available LAR to provide surrogate informed consent
- Age less than or equal to 54
- Patient on comfort care only
- Predicted to have less than 24 hours survival
- Patient enrolled in an investigational drug or device study at the time of enrollment
- Investigator discretion that patient is not suitable for the study
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- SINGLE_GROUP
- Arm && Interventions
Group Intervention Description SafeBreak Vascular Intervention SafeBreak Vascular Every study participant will have SafeBreak Vasculars installed in each of their IV lines.
- Primary Outcome Measures
Name Time Method IV dislodgement rate Up to 7 days Compare the study intervention's rate of IV dislodgement to the clinical literature's prospective observational studies that have a mean dislodgement of 17.5% and median of 9.2%.
- Secondary Outcome Measures
Name Time Method IV restarts Up to 7 days. Comparison of the study's intervention to the clinical literature's rates for IV failure, which is 46% of IVs fail before the end of their intended use
Device related adverse events Up to 7 days Collection of any adverse events related to SafeBreak Vascular
Trial Locations
- Locations (1)
Barnes-Jewish Hospital
🇺🇸Saint Louis, Missouri, United States