Skip to main content
Clinical Trials/NCT04381507
NCT04381507TerminatedNot Applicable

CONFORM-TAA - a Post-market Clinical Follow-up Study in Patients with a Descending Thoracic Aortic Aneurysm or Penetrating Aortic Ulcer (PAU) Treated with the E-nya Thoracic Stent Graft System

JOTEC GmbH1 site in 1 country1 target enrollmentStarted: June 12, 2020Last updated:
Conditions

Trial Snapshot

Phase
Not Applicable
Status
Terminated
Sponsor
JOTEC GmbH
Enrollment
1
Locations
1
Primary Endpoint
Mortality

Study Overview

Brief Summary

The CONFORM-TAA post-market clinical follow-up study is undertaken to evaluate the prevention of death related to a descending aortic aneurysm or PAU when treated by the E-nya Thoracic Stent Graft System. The secondary objective is to evaluate the safety and clinical performance of the device.

Detailed Description

In this study, patients who receive an E-nya Thoracic Stent Graft for the endovascular treatment of descending thoracic aortic aneurysm or PAU will be observed. The E-nya Thoracic Stent Graft will be implanted in accordance with the instructions for use of the E-nya Thoracic Stent Graft System and at the discretion of the treating physician.

Participating physicians will be asked to provide their observations collected during routine standard of care for patients he/she had decided to treat with the E-nya Thoracic Stent Graft System. Informed consent of the patients will be obtained to allow the use of their clinical records for the purpose of this observational study before data are being collected.

The period of data collection for each patient will be approximately 60 months from the intervention. Source document verification will be performed on 100% of patients; data from all visits will be reviewed and verified against existing source documents. Complete DICOM image files of the CT scans will be sent to the CoreLab for independent evaluation.

Study Design

Study Type
Observational
Observational Model
Cohort
Time Perspective
Prospective

Eligibility Criteria

Ages
18 Years to 85 Years (Adult, Older Adult)
Sex
All
Accepts Healthy Volunteers
No

Inclusion Criteria

  • Age between 18 and 85 years
  • A fusiform focal TAA ≥ 55 mm for females, ≥ 60 mm for males, or ≥ 2 times the diameter of the non-aneurysmal thoracic aorta and/or focal saccular TAA or penetrating aortic ulcer (PAU)
  • Suitable proximal and distal landing zone in the native aorta
  • Landing zone of the proximal edge of the fabric distal to the left carotid artery
  • Landing zone of the distal edge of the fabric proximal to the celiac trunk
  • Proximal and distal non-aneurysmal neck diameter between 18 and 42 mm
  • Proximal non-aneurysmal neck length ≥ 20 mm
  • Distal non-aneurysmal neck length ≥ 20 mm
  • Thoracic aortic lesion confirmed by thin sliced (≤ 1 mm) CTA with optional three-dimensional reconstruction obtained within 3 months prior to the implant procedure
  • Patient is able and willing to undergo follow-up imaging and examinations prior to discharge from the hospital, at 30 days and 12 months, and annually thereafter until 5 years follow-up
  • Patient understands and has signed the Informed Consent Form prior to intervention

Exclusion Criteria

  • Female of child bearing potential, breast feeding
  • Access vessels not suitable for endovascular treatment
  • Significant circular thrombi or calcification in proximal or distal landing zones
  • Genetic connective tissue diseases (e.g. Marfan syndrome or Ehler-Danlos Syndrome)
  • Allergies against materials necessary for endovascular repair (e.g. contrast media, heparin, materials of the stent graft)
  • Systemic or local infections
  • eGFR < 45 ml/min/1.73m2 before the intervention
  • Mycotic aneurysm
  • Myocardial infarction or cerebrovascular accident < 3 months ago
  • Patient has specified disease of the thoracic aorta which is not included in the registry, for example: aortic dissection, intramural hematoma, traumatic injury or transection, aortic false aneurysm, (contained) ruptured aneurysm
  • Patients who are planned to be treated with a chimney in the left subclavian artery
  • Treatment of a thoracoabdominal or infrarenal aneurysm at the time of implant
  • Previous stent and/or stent graft or previous surgical repair of descending thoracic aorta
  • Patient had or planned to have a major surgical or interventional procedure within 30 days before or 30 days after the planned implantation of the E-nya Stent Graft System(s). This exclusion does not include planned procedures that are needed for the safe and effective placement of the stent graft (ie, carotid/subclavian transposition, carotid/subclavian bypass procedure)
  • Other medical condition that may cause the patient to be non-compliant with the protocol, confound the results, or is associated with a limited life expectancy (i.e. heart failure, active malignancy (progressive, stable or partial remission))
  • Simultaneously participating in another clinical trial
  • NYHA class IV

Outcomes

Primary Outcomes

Mortality

Time Frame: 30-day

Rate of all-cause mortality

Secondary Outcomes

  • Conversion to open surgery(24hours, 30-day, 12, 24, 36, and 60 months)
  • Endoleak Type Ia(prior to discharge/30-day, 12, 24, 36, and 60 months)
  • Endoleak Type Ib(prior to discharge/30-day, 12, 24, 36, and 60 months)
  • Endoleak Type II(prior to discharge/30-day, 12, 24, 36, and 60 months)
  • Rupture(30-day, 12, 24, 36, 60 months)
  • Primary clinical success(prior to discharge/30-day, 12, 24, 36, and 60 months)
  • Clinical success(prior to discharge/30-day, 12, 24, 36, and 60 months)
  • Primary technical success(24 hours)
  • Technical success(24 hours)
  • Reintervention(30-day, 12, 24, 36, and 60 months)
  • Aneurysm size(12, 24, 36, and 60 months)
  • Cerebrovascular event(30-day, 12, 24, 36, and 60 months)
  • Migration(prior to discharge/30-day, 12, 24, 36, and 60 months)
  • Mortality(30-day, 12, 24, 36, 60 months)
  • Endoleak Type IV(prior to discharge/30-day, 12, 24, 36, and 60 months)
  • Major adverse event(s)(30-day, 12, 24, 36, and 60 months)
  • Endoleak Type III(prior to discharge/30-day, 12, 24, 36, and 60 months)
  • Permanent paraplegia(30-day, 12, 24, 36, and 60 months)
  • Permanent paraparesis(30-day, 12, 24, 36, and 60 months)
  • Integrity(prior to discharge/30-day, 12, 24, 36, and 60 months)
  • Dislodgement(prior to discharge/30-day, 12, 24, 36, and 60 months)
  • Infolding(prior to discharge/30-day, 12, 24, 36, and 60 months)

Investigators