Efficacy of Blanketrol III and Arctic Sun in Cooling Non-shivering Humans

University of Manitoba2 个研究点 分布在 1 个国家目标入组 12 人开始时间: 2020年3月1日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
12
试验地点
2
主要终点
Skin heat exchange

研究概览

简要总结

This study compares three medical cooling devices for effectiveness of cooling the core of non-shivering subjects. This relates to common protocols to cool patients experiencing myocardial infarctions or stokes.

详细描述

Body core and brain cooling by 1-3°C has been used as an adjunct in treatment of stroke and myocardial infarction (MI) patients. Body surface cooling is simple and non-invasive however this process is difficult to implement effectively because skin cooling stimulates shivering heat production, a defense mechanism that protects the body core from cooling. It is sometimes necessary to pharmacologically inhibit shivering in order to cool the body from surface cooling strategies such as application of cold packs or cold water-perfused blankets or pads.

Two commercial cooling units that are currently used are: 1) the Blanketrol III with standard Gel Pad (Gentherm Inc.); and 2) the Arctic Sun with hydrogel coated pads (Arctic Sun). A new cooling attachment system has been developed for the Blanketrol III. The new Kool Kit includes a head wrap, and a vest and lower body blanket. The effectiveness of the Kool Kit has not been quantified or compared to the current two systems mentioned above.

Since shivering heat production attenuates core cooling, it must be eliminated in clinical or research settings where core cooling effectiveness is the main dependent variable. Previous studies from this laboratory (Laboratory for Exercise and Environmental Medicine) have used a human model for severe hypothermia which meperidine (Demerol) is infused to inhibit shivering. Under these conditions the primary core cooling capacity of a cooling system can be determined.

The investigators intend to evaluate the Tco cooling rate, in hospital treatment conditions, [ambient temperature (Ta) = 22ºC] during surface cooling of non-shivering subjects with either: 1) Blanketrol with Gel Pads; 2) Blanketrol with Kool Kit; or 3) an Arctic Sun cooling system.

The purpose is to compare the core cooling rates induced by three commercial surface cooling devices in non-shivering subjects. Because of the increased total surface area of the cooling pads and blankets, the investigators hypothesize that the Blanketrol with Kool Kit will have a greater rate of core cooling than the other two cooling systems tested.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Treatment
盲法
None

入排标准

年龄范围
18 Years 至 45 Years(Adult)
性别
All
接受健康志愿者

入选标准

  • Healthy adults, who answer "No" to all the questions on a Get Active Questionnaire (GAQ) and the "Medical Screening Questionnaire for Hypothermia Study'.
  • Participants will also be interviewed to determine that they are free of any cardio- respiratory disease, Raynaud's Syndrome, or any other conditions that can be aggravated by cold exposure.
  • Also, the study will include women of child bearing potential willing to use highly effective birth control during the trial. All women of child bearing potential will complete the "Research participant information for female subjects" and undergo a pregnancy test at screening.

排除标准

  • Anyone who gives a positive answer to any question in the GAQ or Medical Screening Questionnaire for Hypothermia Study, or who has any cardio-respiratory disease, renal dysfunction, Raynaud's Syndrome, past recreational drug use or any other conditions that can be aggravated by cold exposure will be excluded.
  • Subjects contraindicated for meperidine administration or who have any of the conditions listed below will also be excluded from the study.
  • hypersensitivity to meperidine or any ingredients in the formulation
  • known or suspected mechanical GI obstruction or any diseases/conditions that affect bowel transit
  • suspected surgical abdomen (i.e. acute appendicitis or pancreatitis)
  • severe CNS depression, head injury, increased cerebrospinal or intracranial pressure
  • convulsive disorder, delirium tremens
  • hypothyroidism
  • prostatic hypertrophy or urethral stricture
  • sickle cell anemia
  • Addison's disease
  • Pheochromocitoma
  • Known sensitivity or intolerance to the drug metoclopramide
  • Use of medications that might interact negatively with meperidine. Examples should include MAO inhibitors at the time of screening or within 14 days of screening; CNS depressants; phenytoin; cimetidine; ritonavir; aclyclovir; skeletal muscle relaxants.
  • pregnant women, breastfeeding women, and women planning on becoming pregnant

结局指标

主要结局

Skin heat exchange

时间窗: 120 minutes

Body surface heat flux

Esophageal temperature cooling rate

时间窗: 120 minutes