跳至主要内容
临床试验/NCT04074980
NCT04074980已完成不适用

Post CE Study: Evaluation of the Accuracy of INR Results From Patients Taking Phenprocoumon as Anticoagulant When Measured Using the Lumiradx INR Test

LumiraDx UK Limited1 个研究点 分布在 1 个国家目标入组 102 人开始时间: 2019年5月9日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
102
试验地点
1
主要终点
To determine the agreement of INR measurements of capillary whole blood as assessed by the LumiraDx POC INR test to INR measurements of venous plasma as assessed by the IL ACL ELITE PRO reference method.

研究概览

简要总结

This study is to evaluate the INR levels in patients taking phenprocoumon using the LumiraDx Point of Care INR test and comparing the results to the laboratory reference method , the ACL Elite Pro. Comparison will also be made with the Roche Coaguchek point of care INR test (point of care reference method). Performance will be demonstrated across a subject population of adult males and females in receipt of Phenprocoumon VKA therapy. This population is representative of the indications for use of the CE marked product. Operators will be trained clinical site staff. Venous plasma and capillary whole blood samples will be obtained from each subject in order to assess accuracy of the the LumiraDx Platform when compared to the reference devices (ACL Elite and Coaguchek Pro).

详细描述

Vitamin K antagonists (VKA) are a group of substances that reduce blood clotting by reducing the action of vitamin K. They are used in anticoagulant medications in the prevention of thrombosis. The drugs work by depleting the active form of the vitamin by inhibiting Vitamin K epoxide reductase and thus the recycling of inactive vitamin K back to the active reduced form. The most widely prescribed VKAs are Coumarins. This group is composed of several different structurally similar molecules and includes warfarin (Coumadin), acenocoumerol and phenprocoumon.

Within the EU there are country to country differences in preferential VKA prescription (1). In Germany and surrounding countries, Phenprocoumon (marketed as Marcoumar or Marcumar) is the most commonly prescribed. The majority of data supporting the development and validation of the LumriaDx INR test has used samples form patients taking Warfarin. The aim of this study is to evaluate the INR levels in patients taking phenprocoumon using the LumiraDx Point of Care INR test and comparing the results to the laboratory reference method , the ACL Elite Pro. Comparison will also be made with the Roche Coaguchek point of care INR test (point of care reference method).

This study is an observational, cross-sectional study with 100 patients in total. Devices used in this study are CE marked for INR measurement by healthcare professionals. This study will be conducted according to the principles of Good Clinical Practice (ICH GCP E6(R2)), International Conference on Harmonization Guidelines as applicable to devices, applicable geographical regulations and institutional research policies and procedures.

For part 1 of the study (25 patients), one venous whole blood draw will be performed (~9mL) into two anti-coagulated collection tubes (2 x 4.5mL 3.2% trisodium citrate). These tubes will then by processed to plasma by centrifugation, stored in Eppendorf tubes, labelled and frozen at -80ºC . These will then be transferred to the LumriaDx laboratory in Scotland, UK, for subsequent measurement of INR with the reference device IL ACL Elite Pro Coagulation Analyzer (ACL) and with the LumiraDx POC INR test.

For part 2 of the study (75 patients), one finger-stick sample of capillary blood (~10µl/sample) from each subject, will be applied directly to a unique test strip for immediate measurement of INR on the LumiraDx Instrument. One further fingerstick sample is then obtained (~10µl/sample) from a separate finger for immediate measurement of INR on the Coaguchek PRO. Post-testing, samples will be disposed securely. From each patient, one venous whole blood draw will also be performed (~9mL) into two anti-coagulated collection tubes (2 x 4.5mL 3.2% trisodium citrate). These tubes will then by processed to plasma by centrifugation, stored in Eppendorf tubes, labelled and frozen at -80ºC. These will then be transferred to the LumiraDx laboratory in Scotland, UK, for subsequent measurement of INR with the reference device IL ACL Elite Pro Coagulation Analyzer (ACL) and with the LumiraDx POC INR test.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Persons >18 years of age
  • Willing and able to provide written informed consent and comply with study procedures
  • Currently prescribed Phenprocoumon VKA Therapy
  • Deemed medically appropriate for study participation by the Investigator

排除标准

  • Subject has previously participated in this part of the study
  • Subject is within 4 weeks of first prescription of Phenprocoumon
  • Subject is taking a DOAC, or non-Phenprocoumon VKA therapy
  • Confirmed or suspected pregnancy
  • Vulnerable populations deemed inappropriate for study by the Investigator
  • Deemed medically inappropriate for study by the Investigator (i.e. patients with a known inherited [e.g. haemophilia or von Willebrand's disease] or acquired [e.g. liver cirrhosis] condition likely to be associated with a coagulopathy; or patients receiving non-VKA anticoagulant medications)

研究组 & 干预措施

Venous only

Experimental

One venous whole blood draw will be performed into two anti-coagulated collection tubes

干预措施: Venous blood draw (Diagnostic Test)

Fingerstick and Venous

Experimental

One finger-stick sample of capillary blood (~10µl/sample) from each subject, will be applied directly to a unique test strip for immediate measurement of INR on the LumiraDx Instrument. One further fingerstick sample is then obtained (~10µl/sample) from a separate finger for immediate measurement of INR on the Coaguchek PRO. From each patient, one venous whole blood draw will also be performed into two anti-coagulated collection tubes.

干预措施: Venous and Fingerstick blood draws (Diagnostic Test)

结局指标

主要结局

To determine the agreement of INR measurements of capillary whole blood as assessed by the LumiraDx POC INR test to INR measurements of venous plasma as assessed by the IL ACL ELITE PRO reference method.

时间窗: 6 weeks

Measurement of INR using Venous blood

次要结局

  • To determine the agreement of INR measurements of capillary whole blood as assessed by the LumiraDx POC INR test to INR measurements of capillary blood as assessed by Coaguchek Pro II.(6 weeks)

研究者

发起方
LumiraDx UK Limited
申办方类型
Industry
责任方
Sponsor

研究点 (1)

Loading locations...