Validation Study in Patients of a Non-invasive Blood Pressure Monitor for Perioperative Use

Academisch Medisch Centrum - Universiteit van Amsterdam (AMC-UvA)1 个研究点 分布在 1 个国家目标入组 100 人开始时间: 2017年3月1日最近更新:
适应症

试验速览

阶段
不适用
入组人数
100
试验地点
1
主要终点
validation systole, diastole, mean ABP and pulse pressure

研究概览

简要总结

Collecting all available data (waveforms, beat to beat data, status data) generated by a non invasive blood pressure monitor on each hand and compare this to the actual data obtained by intra arterial (radial) monitoring.

This to see if the non invasive bloodpressure monitor can be validated for intraoperative use.

详细描述

Although a blood pressure monitoring device could be evaluated by the comparing systolic, diastolic and mean arterial pressure to a reference pressure, it is the specific aim to collect all available data (waveforms, beat to beat data, status data) generated by the blood pressure monitor.

After data analysis and comparison to patient's actual intra-arterial waveform, the outcome will be evaluated by Association for the Advancement of Medical Instrumentation (AAMI) standards to see if the non invasive bloodpressure monitor can be validated for intraoperative use.

The study is targeted at patients already monitored with a system that accurately measures and stores the intra-arterial waveform, such as the FloTrac system in the EV1000 monitor.

This study is a substudy of a feasability protocol evaluating a new algorithm for the Clearsight system.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Patients whose arterial blood pressure is measured invasively with technology that allows for the storage of the intra-arterial waveform with sufficient quality and resolution (Flotrac).
  • Patients older than 18, for elective general surgery.

排除标准

  • Patient conditions that are expected to potentially and significantly affect the transfer of pressure between radial artery and the finger, e.g. due to recent finger fractures

结局指标

主要结局

validation systole, diastole, mean ABP and pulse pressure

时间窗: within 1 hour

accuracy and precision of invasive blood pressure curve vs non-invasive blood pressure curves(ipsilateral and contralateral)

implementation of software alterations to improve systole, diastole, mean ABP accuracy and pulse pressure

时间窗: time during surgery

Comparison of the non-invasive blood pressure curve after software alterations to the reference and invasive arterial blood pressure curve

次要结局

未报告次要终点

研究者

申办方类型
Other
责任方
Principal Investigator
主要研究者

D.P.Veelo

Principal investigator

Academisch Medisch Centrum - Universiteit van Amsterdam (AMC-UvA)