Retrospective, Observer-blinded, Pivotal Study to Evaluate the Effectiveness of the Investigational Device (Lunit INSIGHT CXR) in Detection of Abnormal Findings on Chest Radiographs

Lunit Inc.1 个研究点 分布在 1 个国家目标入组 1,111 人开始时间: 2021年6月2日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
Lunit Inc.
入组人数
1,111
试验地点
1
主要终点
Area Under the Receiver Operating Characteristic Curve (ROC AUC) of the investigational device in detection of overall target radiologic findings

研究概览

简要总结

This study aims to evaluate the effectiveness of the investigational device (Lunit INSIGHT CXR) in the detection of chest abnormalities.

详细描述

In the study, the standalone performance of Lunit INSIGHT CXR was primarily assessed by comparing the analysis results from the investigational device and reference standards in the detection of 10 abnormal radiologic findings - atelectasis, calcification, cardiomegaly, consolidation, fibrosis, mediastinal widening, nodule/mass, pleural effusion, pneumoperitoneum, and pneumothorax.

In addition, to investigate the effect of Lunit INSIGHT CXR assistance on the interpreting physician's interpreting performance, a multi-reader, multi-case (MRMC) study was conducted.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Retrospective

入排标准

年龄范围
14 Years 至 —(Child, Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Chest radiographs of aged 14 years or older
  • Conventional PA or AP chest radiographs
  • Chest radiographs with confirmed radiology reports

排除标准

  • Chest radiographs taken in postures other than PA and AP e.g.) decubitus, oblique, lordotic, or lateral view
  • Invert-grayscale chest radiographs
  • Image quality not suitable for interpretation due to incorrect patient positioning, inadequate image collimation, severely damaged lung presence of external materials such as jewelry

结局指标

主要结局

Area Under the Receiver Operating Characteristic Curve (ROC AUC) of the investigational device in detection of overall target radiologic findings

时间窗: Through study completion, an average of 6 months

To evaluate the effectiveness of the investigational device (Lunit INSIGHT CXR) in the detection of chest abnormalities, Area Under the Receiver Operating Characteristic Curve (ROC AUC) of the investigational device in detection of overall target radiologic findings (detection of any abnormal findings among ten target radiologic findings) will be measured