Calibration of the Cloud DX Pulsewave Health Monitor (PAD-2A) Oscillometric Wrist Cuff Home Blood Pressure Monitor, According to the AAMI/ESH/ISO Universal Standard (ISO 81060-2: AMD_2020) (CCV-4)

Cardiovascular Research New Brunswick1 个研究点 分布在 1 个国家目标入组 200 人开始时间: 2021年6月15日最近更新:
适应症

试验速览

阶段
不适用
状态
招募中
发起方
入组人数
200
试验地点
1
主要终点
Diastolic Pressure

研究概览

简要总结

The objective of this study is to calibrate the blood pressure measurements of the Pulsewave Health Monitor (PAD-2A) device to be within 5±8 mmHg of the average of the dual-observer measurements via an aneroid sphygmomanometer (reference device), as per the methodology described in the ISO 81060-2: AMD_2020 protocol.

详细描述

The plan for this study is to calibrate the blood pressure measurements of the PAD-2A device with the dual-observer blood pressure measurements via an aneroid sphygmomanometer (reference device), as per the ISO 81060-2: AMD_2020 protocol. The data will be collected and analyzed in order to calibrate the PAD-2A device blood pressure measurements to be within 5±8 mmHg of the average of the dual-observer auscultatory blood pressure measurements by adjusting the device algorithm.

研究设计

研究类型
Observational
观察模型
Case Only
时间视角
Prospective

入排标准

年龄范围
19 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Adults ≥ 19 years of age (minimum 30% male/female)
  • Wrist circumference between 13.5 cm. - 23 cm. (5.3 in. - 9.1 in.):
  • Specifically for this test device, the wrist size of the consented participants will be distributed accordingly for a Gaussian distribution:
  • Lowest Octile: ≥ 13.5cm to < 14.7cm Lower Quarter: ≥ 13.5cm to < 15.9cm Lower-Middle Quarter: ≥ 15.9cm to < 18.3cm Upper-Middle Quarter: ≥ 18.3cm to < 20.7cm Upper Quarter: ≥ 20.7cm to ≤ 23.0cm Highest Octile: ≥ 21.9cm to ≤ 23.0cm
  • ● Willing to volunteer to participate and to sign the study specific informed consent form

排除标准

  • Wrist circumference less than 13.5 cm. (5.3 in.) or greater than 23 cm. (9.1 in)
  • Hand or body tremors
  • Irregular heart rhythm (bigeminy, trigeminy, isolated ventricular premature beat (VPB), atrial fibrillation)
  • Korotkoff sound K5 not audible
  • A musculoskeletal disorder that prevents a non-invasive device to be inflated/deflated on the arm
  • Unwilling to volunteer to participate and to sign the study specific informed consent form

结局指标

主要结局

Diastolic Pressure

时间窗: From the start of manual inflation of the upper arm blood pressure cuff to the end of (manual) deflation of the upper arm blood pressure cuff (approximately 30-60 seconds)

Simultaneous auscultatory blood pressure measurements by 2 trained observers

Systolic Pressure

时间窗: From the start of manual inflation of the upper arm blood pressure cuff to the end of (manual) deflation of the upper arm blood pressure cuff (approximately 30-60 seconds)

Simultaneous auscultatory blood pressure measurements by 2 trained observers