Study Overview

Brief Summary

We used a new and easy-to-measure iPad 3D scanner to measure swelling in lower limbs because current technology is not easily available or affordable. We want to see how reliable the measurements are and what might affect how consistent they are.

Detailed Description

Twenty limbs from ten healthy volunteers were scanned using an iPad-based Structure Sensor and software. Initial scans followed standard manufacturer instructions, but high variance rendered data unsuitable for clinical use. To improve accuracy, a standardized scanning protocol was developed, incorporating anatomical calibration, scanning distance standardization, and scanning time control. A 254-mm calf segment was defined using a 3D marker placed on the medial malleolus to ensure consistent volume measurement. The scanning distance was fixed between 50-59 cm to reduce zoom parallax errors, and scans were conducted after 3 PM to minimize diurnal volume fluctuations. Multiple technicians performed repeat scans on the same limb to assess intra-observer and inter-observer scan reliability.

Study Design

Study Type
Observational
Observational Model
Cohort
Time Perspective
Prospective

Eligibility Criteria

Sex
All
Accepts Healthy Volunteers
Yes

Inclusion Criteria

  • The people in this study will be symptom free volunteers.

Exclusion Criteria

  • Pre-existing diseased limbs

Outcomes

Primary Outcomes

Limb Volume

Time Frame: During Procedure

The scanning protocol and the required modifications of "out of the box" instructions for optimal clinical application are described; The 10 inch (254mm) long limb segment scanned must be anatomically constant. We chose the medial malleolus as the lower border of the scan. The scannable limb volume is 254 mm long, measured up from the malleolus.

Secondary Outcomes

No secondary outcomes reported

Investigators

Sponsor
The RANE Center for Venous and Lymphatic Diseases at St. Dominic's Hospital
Sponsor Class
Other
Responsible Party
Principal Investigator
Principal Investigator

Dr. Seshadri Raju

Principal Investigator

The RANE Center for Venous and Lymphatic Diseases at St. Dominic's Hospital

Study Sites (1)

Loading locations...

Similar Trials

Measuring Volume Using a 3D Scanner and... | Clinical Trial