MedPath

Interference Between Surgical Magnetic Drapes and Pacemakers

Not Applicable
Completed
Conditions
Electromagnetic Interference
Interventions
Device: Prototype surgical magnetic drape
Device: Green Covidien Magnetic Drape
Device: Jac-Cell Medic Reusable Magnetic Pad
Device: DeRoyal Magnetic Pad size 20" x 16"
Device: DeRoyal Magnetic Pad size 10" x 16"
Registration Number
NCT02617355
Lead Sponsor
Maisonneuve-Rosemont Hospital
Brief Summary

This study is designed to compare a new prototype surgical magnetic drape made with bottom-isolated ferrite magnets to 4 commercially available surgical magnetic drapes.

Detailed Description

Magnetic fields may potentially interfere with the function of cardiovascular implantable electronic devices. Sterile magnetic drapes used to hold surgical instruments are often placed on the patient's thorax, and they are likely to interfere with the function of these devices.

This study compares the magnetic interference of a new prototype surgical magnetic drape made with bottom-isolated ferrite magnets to the Covidien magnetic drape and 3 other commercially available surgical magnetic drapes.

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
47
Inclusion Criteria
  • Patients with a pacemaker
  • 18 years or older
Exclusion Criteria
  • Patients with implantable cardioverter-defibrillator (ICD)
  • Patients in whom the magnet rate cardiac rhythm is indistinguishable from their baseline rhythm
  • Refusal/Inability to consent
  • Language other than French or English

Study & Design

Study Type
INTERVENTIONAL
Study Design
PARALLEL
Arm && Interventions
GroupInterventionDescription
Prototype vs Covidien magnetic drapeGreen Covidien Magnetic DrapeNew prototype surgical magnetic drape made with bottom-isolated ferrite magnets applied to the patient's thorax vs Covidien magnetic drape
Prototype vs 4 commercial drapesDeRoyal Magnetic Pad size 20" x 16"New prototype surgical magnetic drape made with bottom-isolated ferrite magnets applied to the patient's thorax vs 4 commercially available magnetic drapes: Green Covidien Magnetic Drape Jac-Cell Medic Reusable Magnetic Pad DeRoyal Magnetic Pad size 10" x 16" DeRoyal Magnetic Pad size 20" x 16"
Prototype vs Covidien magnetic drapePrototype surgical magnetic drapeNew prototype surgical magnetic drape made with bottom-isolated ferrite magnets applied to the patient's thorax vs Covidien magnetic drape
Prototype vs 4 commercial drapesGreen Covidien Magnetic DrapeNew prototype surgical magnetic drape made with bottom-isolated ferrite magnets applied to the patient's thorax vs 4 commercially available magnetic drapes: Green Covidien Magnetic Drape Jac-Cell Medic Reusable Magnetic Pad DeRoyal Magnetic Pad size 10" x 16" DeRoyal Magnetic Pad size 20" x 16"
Prototype vs 4 commercial drapesPrototype surgical magnetic drapeNew prototype surgical magnetic drape made with bottom-isolated ferrite magnets applied to the patient's thorax vs 4 commercially available magnetic drapes: Green Covidien Magnetic Drape Jac-Cell Medic Reusable Magnetic Pad DeRoyal Magnetic Pad size 10" x 16" DeRoyal Magnetic Pad size 20" x 16"
Prototype vs 4 commercial drapesJac-Cell Medic Reusable Magnetic PadNew prototype surgical magnetic drape made with bottom-isolated ferrite magnets applied to the patient's thorax vs 4 commercially available magnetic drapes: Green Covidien Magnetic Drape Jac-Cell Medic Reusable Magnetic Pad DeRoyal Magnetic Pad size 10" x 16" DeRoyal Magnetic Pad size 20" x 16"
Prototype vs 4 commercial drapesDeRoyal Magnetic Pad size 10" x 16"New prototype surgical magnetic drape made with bottom-isolated ferrite magnets applied to the patient's thorax vs 4 commercially available magnetic drapes: Green Covidien Magnetic Drape Jac-Cell Medic Reusable Magnetic Pad DeRoyal Magnetic Pad size 10" x 16" DeRoyal Magnetic Pad size 20" x 16"
Primary Outcome Measures
NameTimeMethod
Percentage of magnetic interferenceWithin 2 minutes after placement of surgical magnetic drape to the patient's chest

Percentage of patients experiencing magnetic interference with the application of each surgical magnetic drape.

Magnetic interference defined as a cardiac rhythm change to asynchronous and identical to that produced by a round magnet rhythm over the pacemaker.

Secondary Outcome Measures
NameTimeMethod