Comparison between an automatic sleep-staging program via portable 1-channel electroencephalograph versus manual sleep staging by traditional polysomnography - Comparison between portable 1-channel electroencephalograph versus traditional polysomnography
适应症
试验速览
- 阶段
- 未知
- 状态
- 已完成
- 入组人数
- 41
研究概览
简要总结
Compared with polysomnography, Sleep Scope seemed to overestimate sleep latency by approximately 3 min, but there was no consistent tendency in bias in other sleep parameters. The kappa values ranged from 0.66 to 0.75 for experts' inter-rater polysomnography scores and from 0.62 to 0.67 for Sleep Scope versus polysomnography scores, which indicated sufficient agreement in the determination of sleep stages based on the Landis and Koch criteria.
研究设计
- 研究类型
- Observational
入排标准
- 年龄范围
- 20years-old 至 80years-old(—)
- 性别
- All
入选标准
- 未提供
排除标准
- •Participants had no competent to consent to contract.
研究者
相似试验
已完成
不适用
Ultra Long-Term Sleep Monitoring Using UNEEG™ Medical 24/7 EEG™ SubQSleepNCT04513743UNEEG Medical A/S25
已完成
不适用
Validation of Actigraph Measured Sleep During Acute DiverticulitisAcute DiverticulitisNCT01840878Herlev Hospital13
已完成
不适用
Exploratory study on subjective sleep evaluation and mood state who breakfast skippers regularly eat breakfastHealthy AdultsJPRN-UMIN000041016Kanto Gakuin University194
Unknown
不适用
Verification of Computerized Sleep-wake Analysis in Sleep DisordersSleep-wake AnalysisNCT02412527Chang Gung Memorial Hospital330
已完成
不适用
Efficacy of the Dreem 3S Ambulatory Sleep Monitoring Device for the Evaluation of NarcolepsyNarcolepsy Type 1Sleep DisorderHypersomniaNCT06531876Beacon Biosignals78