First in Human Feasibility Study With ADAPT 3D - ALR for Aortic Leaflet Repair

Not Applicable
Conditions
Aortic Insufficiency
Aortic Stenosis
Interventions
Device: Implantation of ADAPT 3D ALR
Registration Number
NCT04178213
Lead Sponsor
Anteris Technologies Ltd.
Brief Summary

This study will evaluate the safety and performance of the ADAPT 3D - ALR in adult patients requiring replacement of aortic valve. 15 patients in one site in Belgium will all be treated with ADAPT 3D - ALR.

Detailed Description

The purpose of this study is to conduct the initial clinical investigation of single piece ADAPT treated 3D - ALR (Aortic Leaflet Repair) to collect evidence on the device's safety and performance. The study is anticipated to confirm successful clinical safety and clinical performance with significant improvements in clinical hemodynamic performance.

The ADAPT® technology is used to process animal derived tissues to produce implantable tissue prosthetic devices that are compatible with the human body. This technology has been shown to produce reliable, biocompatible and versatile regenerative prosthetic devices capable of being used instead of synthetic products currently used in many soft tissue repair applications.

15 patients with aortic valve insufficiency or stenosis will be enrolled in this single arm single site study.

Follow-up will continue through to 26 weeks.

Recruitment & Eligibility

Status
UNKNOWN
Sex
All
Target Recruitment
15
Inclusion Criteria
  1. Patient is older than 18 years of age and ≤ 85 years.
  2. The subject's aortic annular measurements are ≥ 21 mm to < 27mm as confirmed on pre-op echo.
  3. The subject is a candidate for Aortic Valve Replacement.
  4. The subject has documented moderate or severe Aortic Stenosis and/or Aortic Insufficiency (defined as grade 2, 3 or 4).
  5. The subject is willing and able to comply with specified follow-up requirements and evaluations, including transesophageal echocardiography (TEE) if there are inadequate images by transthoracic echocardiography (TTE) to assess the aortic valve.
  6. Patient has signed the informed consent
Exclusion Criteria
  1. All patients who require emergency surgery (within 24 hours of a presentation to an emergency department) for any reason.
  2. Subject with a pre-existing valve prosthesis in the aortic position.
  3. Patients requiring repair of other cardiac valves will be excluded.
  4. Subject with active endocarditis.
  5. Heavily calcified aortic roots or "porcelain aortas".
  6. Leukopenia with a WBC (white Blood Cells) of less than 3000.
  7. Acute anaemia with a haemoglobin less than 8 g/dL.
  8. Platelet count less than 100,000 platelets/microliter, and if less than 150,000 platelets/microliter, platelet reduction of >10,000 platelets/microliter per day over two consecutive days (Note: Where a platelet count is less than 150,000 platelets/microliter but greater than 100,000 platelets/microliter, platelet count testing is to be conducted over 2 consecutive days. If platelets are reducing over those two consecutive days, the rate of reduction in platelets must be less than 10,000 platelets/microliter per day, otherwise the participant must be excluded).
  9. History of a defined bleeding diathesis or coagulopathy or the subject refuses blood transfusions.
  10. Active infection requiring antibiotic therapy (if temporary illness, subjects may enroll 2-4 weeks after discontinuation of antibiotics).
  11. Subjects in whom transesophageal echocardiography (TEE) is contraindicated.
  12. Low EF < 50 %.
  13. Life expectancy < 1 year, or severe comorbidities, such as cancer, dialysis, or severe COPD.
  14. The subject is an illicit drug user, alcohol abuser, prisoner, institutionalised, or is unable to give informed consent.
  15. The subject is pregnant or lactating (non-pregnancy to be confirmed by pregnancy test for all child bearing potential females).
  16. Recent (within 6 months) cerebrovascular accident (CVA) or a transient ischemic attack (TIA).
  17. Myocardial Infarction (MI) within one month of trial inclusion.
  18. Currently participating in, or have been recently exited from (within 30 days of enrollment in this study), or plan to be enrolled in another clinical

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Arm && Interventions
GroupInterventionDescription
ADAPT 3D ALRImplantation of ADAPT 3D ALRPatients treated with ADAPT 3D ALR
Primary Outcome Measures
NameTimeMethod
rate of major paravalvular leak6 months following implantation

The adverse events consistent with the Objective Performance Criteria (OPC), for flexible heart valves will be compared to those rates defined as acceptable levels in the standard.

derived Effective Orifice Area (EOA) range > 0.9 cm2 (19mm valve) to > 1.6 cm2 (27mm valve)6 months following implantation

Hemodynamic Performance Assessment

rate of major hemorrhage6 months following implantation

The adverse events consistent with the Objective Performance Criteria (OPC), for flexible heart valves will be compared to those rates defined as acceptable levels in the standard.

rate of valve thrombosis6 months following implantation

The adverse events consistent with the Objective Performance Criteria (OPC), for flexible heart valves will be compared to those rates defined as acceptable levels in the standard.

rate of thromboembolism6 months following implantation

The adverse events consistent with the Objective Performance Criteria (OPC), for flexible heart valves will be compared to those rates defined as acceptable levels in the standard.

mean pressure gradient (mmHg) across the valve (less than 20 mmHg)6 months following implantation

Hemodynamic Performance Assessment

rate of endocarditis6 months following implantation

The adverse events consistent with the Objective Performance Criteria (OPC), for flexible heart valves will be compared to those rates defined as acceptable levels in the standard.

Secondary Outcome Measures
NameTimeMethod
Rate of hemorrhage6 months following implantation

The rate will be compared to clinical outcomes for surgically implanted heart valves reported in the literature.

Rate of Atrial Fibrillation 6 months post procedure6 months post procedure

New/post-operative atrial fibrillation - confirmed on ECG after closure till 6 months

NYHA (New York Heart Association) class Improvement Assessment6 months post procedure

Clinically significant improvement (one grade) in the New York Heart Association (NYHA) functional classification. 4 classes of disease: min value Class IV (cardiac disease resulting in inability to carry on any physical activity without discomfort); max value Class I (cardiac disease but without resulting limitations of physical activity)

number of days in hospital post procedure30 days post procedure

Post procedure length of stay defined as the date and time in hours and minutes documented for arrival in the recovery unit to date and time in hours and minutes of discharge in hours and minutes.

hemoglobin assessment6 months post procedure

Hemolysis screen measured by blood test

Aspartate transaminase (AST)6 months post procedure

Hemolysis screen measured by blood test of liver enzyme

Rate of all-caused death6 months following implantation

The rate will be compared to clinical outcomes for surgically implanted heart valves reported in the literature.

Alanine transaminase (ALT)6 months post procedure

Hemolysis screen measured by blood test of liver enzyme

Rate of valve-related reoperation6 months following implantation

The rate will be compared to clinical outcomes for surgically implanted heart valves reported in the literature.

Rate of all-cause reoperation6 months following implantation

The rate will be compared to clinical outcomes for surgically implanted heart valves reported in the literature.

Rate of valve related death6 months following implantation

The rate will be compared to clinical outcomes for surgically implanted heart valves reported in the literature.

number of days in ICU30 days post procedure

Length of stay in the ICU post valve implantation defined as arrival time/date in hours and minutes to transfer to ward time/date in hours and minutes.

Rate of valve explant6 months following implantation

The rate will be compared to clinical outcomes for surgically implanted heart valves reported in the literature.

Rate of Device deficiency6 months following implantation

Device deficiency measured by echo