Can the Diagnostic Accuracy of Newborn Eye Screening for Congenital Cataract be Improved With Digital Dual Light Source Imaging? The Digital Imaging vs Ophthalmoscopy (DIvO)Study.

Cambridge University Hospitals NHS Foundation Trust0 个研究点目标入组 140,000 人开始时间: 2022年6月1日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
尚未招募
入组人数
140,000
主要终点
Sensitivity of the screening evaluation of the Neocam (intervention) test

研究概览

简要总结

Cataract (cloudiness of the lens) is the major cause of avoidable child blindness in the world and affects 1 in 3000 UK infants. Screening may fail to detect a cataract in an affected child (false-negative) or mistakenly suggest there is a cataract (false positive) triggering urgent unnecessary referral.

Screening is currently undertaken using an ophthalmoscope into the eye to assess the reddish reflected light (red-reflex). This study aims to test if screening using a new hand-held digital imaging device (Neocam) is more accurate than the ophthalmoscope for newborn eye screening.

详细描述

Cataract (cloudiness of the lens) is the major cause of avoidable child blindness in the world and affects 1 in 3000 UK infants. Cataracts are present in both eyes in more than half of the babies affected. Surgery is required by 8 weeks of age to prevent permanent visual impairment at this critical time of vision and brain development. All UK babies are examined (screened) for cataract twice within the first 8 weeks of life but late diagnosis continues to be a problem, causing avoidable visual impairment in some affected children.

Screening may fail to detect a cataract in an affected child (false-negative) or mistakenly suggest there is a cataract (false positive) triggering urgent unnecessary referral, parental anxiety and wasted NHS resources. Screening is currently performed by midwives and doctors shining a bright white light torch (an ophthalmoscope) into the eye to assess the reddish reflected light (red-reflex), similar to "red eye" seen in flash photos. Cataract causes a dark shadow on the red-reflex but the test can be difficult because bright light causes the pupils to constrict and the babies to forcefully shut their eyes. Assessment is particularly difficult in ethnic minority infants since eye pigmentation affects the hue and brightness of the red-reflex.

This study aims to test if screening using a new hand-held digital imaging device (Neocam) is more accurate than the ophthalmoscope for newborn eye screening. Neocam painlessly images the eye's reflection firstly to infrared light and then to a brief green light flash. Its infrared light causes no pupil constriction or avoidance response and the reflection is bright and consistent regardless of the baby's ethnicity. Its brief green flash and immediate imaging allows a photo of the red-reflex to be captured before the pupil has time to constrict.

To compare the accuracy of both tests, the investigators seek to enrol 140,000 newborn babies in a two year study period. All babies will have both the current ophthalmoscope screening test and additional Neocam imaging. If either test is potentially abnormal, the baby will be referred for specialist examination.

A more accurate screening test could prevent life-long disability and reduce costs to the NHS and society. This study will allow a future estimation of what these savings might be and whether changing to a digital imaging screening service might be justified. The result may have an impact on eye screening world-wide to prevent childhood blindness from cataract.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Screening
盲法
None

入排标准

年龄范围
— 至 72 Hours(Child)
性别
All
接受健康志愿者

入选标准

  • All newborn babies having the newborn physical examination

排除标准

  • 未提供

研究组 & 干预措施

Babies undergoing newborn eye screening

Other

Babies having newborn eye screening with the intervention (Digital Imaging) technique in addition to the standard ophthalmoscopic technique.

干预措施: Neocam (Device)

结局指标

主要结局

Sensitivity of the screening evaluation of the Neocam (intervention) test

时间窗: 2 years

The proportion of patients correctly identified as having cataract by a positive Neocam screening test result. Where a higher value indicates better sensitivity of the screening test for presence of cataract.

Specificity of the screening evaluation of the Neocam (intervention) test

时间窗: 2 years

The proportion of patients correctly identified as not having cataract by a negative Neocam screening test result. Where a higher value indicates better specificity of the screening test for absence of cataract.

Sensitivity of the screening evaluation of the standard test

时间窗: 2 years

The proportion of patients correctly identified as having cataract by a positive standard screening test result. Where a higher value indicates better sensitivity of the screening test for presence of cataract.

Specificity of the screening evaluation of the standard test

时间窗: 2 years

The proportion of patients correctly identified as not having a cataract by a negative standard screening result. Where a higher value indicates better specificity of the screening test for absence of cataract.

次要结局

  • Usability feedback(2 years)

研究者

申办方类型
Other
责任方
Principal Investigator
主要研究者

Dr Louise Allen

Chief Investigator

Cambridge University Hospitals NHS Foundation Trust