StatStrip Glucose/Ketone Meter System Ketone Evaluation
Trial Snapshot
- Phase
- Not Applicable
- Status
- Completed
- Sponsor
- Nova Biomedical
- Enrollment
- 173
- Locations
- 3
- Primary Endpoint
- Analytical verification of StatStrip glucose/ketone meter system - ketone comparison
Study Overview
Brief Summary
The Nova StatStrip Glucose / Ketone Meter System is a fast, simple, whole blood system that can be used to quantify both glucose and ketones using two distinct test strips. This Protocol will describe an evaluation of the system's Ketone performance only. Ketone testing methods that quantify beta-hydroxybutyrate (β-OHB), the predominant ketone body, are used for diagnosing and monitoring ketoacidosis, a life-threatening complication of hyperglycemia. The performance of the StatStrip Ketone test strip using the Nova StatStrip Glucose / Ketone Hospital Meter System on venous and capillary whole blood will be assessed.
Detailed Description
The primary objectives of this study are:
- To assess the performance of the StatStrip Ketone assay in the hands of CLIA-Waived Point-of-Care users in at least three (3) distinct Point-of-Care clinical settings on venous, and capillary blood and compare the performance characteristics to a traceable laboratory reference method.
- To assess the Ease of Use of the StatStrip Glucose / Ketone Meter System in the hands of the intended CLIA-Waived Point-of-Care users. CLIA-Waived operators will be provided with all package-insert sheets, a Quick Reference Guide and Instructions for Use (IFU) Manual. No training, coaching, or prompting will be provided to the CLIA-Waived POC Users other than clarifying the protocol.
Study Design
- Study Type
- Observational
- Observational Model
- Cohort
- Time Perspective
- Prospective
Eligibility Criteria
- Ages
- 18 Years to — (Adult, Older Adult)
- Sex
- All
- Accepts Healthy Volunteers
- Yes
Inclusion Criteria
- •Adult subjects (≥ 18 years old) with diabetes or suspected ketonemia.
- •Adult subjects (≥ 18 years old) that are healthy.
- •Subjects willing and able to consent to participating in the study.
- •Subjects whose pre-screen ketone value, if performed, is deemed valuable to the study.
Exclusion Criteria
- •Subjects unable to consent to participating in the study.
- •Subjects possessing a cognitive disorder or other condition, which, in the opinion of the investigator, would put the person at risk or seriously compromise the integrity of the study.
Outcomes
Primary Outcomes
Analytical verification of StatStrip glucose/ketone meter system - ketone comparison
Time Frame: 20 days
StatStrip ketone meter system, a point of care testing instrument is as effective as a reference laboratory method for ketone results in mmol/L
Secondary Outcomes
No secondary outcomes reported