Multi Parameter Vital Signs Monitoring by the Corsano CardioWatch 287-2 Validation Study

Corsano Health B.V.1 个研究点 分布在 1 个国家目标入组 100 人开始时间: 2022年10月19日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
100
试验地点
1
主要终点
RMSE of Heart Rate

研究概览

简要总结

Today, continuous monitoring of vital signs remains a challenge since it generally requires the patient to be connected to multiple wired sensors, which restricts patient mobility in the intra-mural setting and complicates home monitoring in the extra-mural setting. Wearable devices on the wrist, although emerging, are often not clinically validated or limited to the monitoring of one or two vital signs.

This study aims to validate the Corsano CardioWatch 287-2 for the continuous monitoring of heart rate at ≤ 4 bpm root mean squared error (RMSE); interbeat intervals at ≤ 50 ms RMSE; breathing rate at ≤ 2 brpm RMSE; and peripheral oxygen saturation at ≤ 3 percentage point RMSE. Also, this study aims to validate the Corsano CardioWatch 287-2 for the measurement of non-invasive blood pressure according to ISO 81060-2:2018.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

性别
All
接受健康志愿者

入选标准

  • 18 years old;
  • undergoing invasive monitoring, like coronary angiography;
  • able to provide consent.

排除标准

  • who cannot wear the Corsano CardioWatch 287 due to reasons such as allergic reactions, wounds, amputations etc.;
  • unable or not willing to sign informed consent;
  • with significant mental or cognitive impairment;
  • who do not have a suitable entry site for the invasive arterial line
  • who do not comply to criteria established by ISO 81060-2:2018.

结局指标

主要结局

RMSE of Heart Rate

时间窗: 30 minutes

Root-mean-squared error between PPG measured and gold standard measured heart rate

RMSE of Respiratory Rate

时间窗: 30 minutes

Root-mean-squared error between PPG measured and gold standard measured respiratory rate

RMSE of Oxygen Saturation

时间窗: 30 minutes

Root-mean-squared error between PPG measured and gold standard measured oxygen saturation

RMSE of Blood Pressure

时间窗: 30 minutes

Root-mean-squared error between PPG measured and gold standard measured blood pressure (systolic blood pressure and diastolic blood pressure)