FreeStyle Libre Flash Glucose Monitoring System Accuracy Study
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 50
- 试验地点
- 4
- 主要终点
- System Performance
研究概览
简要总结
This is a non-randomized, single arm, multi-center, prospective, non-significant risk study to evaluate the FreeStyle Libre Flash Glucose Monitoring System.
详细描述
Up to 60 subjects will be enrolled at up to six clinical research sites in the United States. Subjects will wear two Sensors of Three unique sensor lots. Each Sensor will have a paired Reader that will be given to the subject. All Readers will be masked during the study (i.e. subjects will not be able to view glucose results obtained from the Sensor on the Reader screen). Subjects will be asked to perform at least 8 capillary Blood Glucose (BG) tests per day using the primary Reader. Interstitial glucose readings from each Sensor will be obtained with the corresponding Readers immediately following each BG test. Subjects will follow the same BG testing schedule during in-clinic visits. Subjects will be instructed to report any problems with the device. Subjects will make seven (7) scheduled visits to the clinical study site, including the Enrollment/Screening Visit (Visit 1). Subjects will have four (4) in-clinic visits during which intravenous blood draws and YSI reference testing will occur.
研究设计
- 研究类型
- Observational
- 观察模型
- Cohort
- 时间视角
- Prospective
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- 未提供
排除标准
- 未提供
结局指标
主要结局
System Performance
时间窗: Approximately 32 hours
System performance will be characterized with respect to YSI reference venous plasma sample measurements.
System Related adverse device effects
时间窗: Up to 42 days
System will be characterized by Adverse Device Effects and Serious Adverse Device Effects experienced by study participants.
次要结局
未报告次要终点