Rotational Stability of a Single-piece Toric Acrylic Intraocular Lens: a Pilot Study

Prim. Prof. Dr. Oliver Findl, MBA0 个研究点目标入组 30 人开始时间: 2011年3月最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
30
主要终点
Rotational stability of the investigated toric lens

研究概览

简要总结

Abstract Purpose: To evaluate the visual performance and rotational stability of the Tecnis Toric one-piece IOL (AMO, USA) during the first 3 post-operative months Design: Prospective single center study Setting: VIROS - Vienna Institute for Research in Ocular Surgery, a Karl Landsteiner Institute, Hanusch Hospitel, Vienna, Austria Methods: In this prospective study patients with age-related cataract and corneal astigmatism of 1.0 to 3.0 D measured with the IOL-Master 500 (Carl Zeiss Meditec AG, Germany) were included. Pre-operatively, rotating Scheimpflug scans (Pentacam HR, Oculus, Germany) were performed and the cornea was marked in the sitting position at the slit lamp. Patients recieved a single-piece toric hydrophobic acrylic IOL (Tecnis Toric, AMO, USA). Immediately and 3 months after surgery retroillumination photographs were taken to assess the rotational stability of the IOL. Additionally, Autorefraction (Topcon, USA), subjective refraction, uncorrected and distance corrected visual acuity, keratometry, Scheimpflug and ocular wavefront (WASCA, Carl Zeiss Meditec AG, Germany) measurements were performed at the 3 months follow-up.

详细描述

Whereas first designs of toric intraocular lenses (toric IOLs) in the early 1990's showed an IOL rotation of more than 30° in one fifth of the patients, modern toric IOLs typically show a mean absolute rotation of 3° to 5°, which would result in a loss of about 10% to 15% of the astigmatism reducing effect of the toric IOL.

Rotational stability of a toric IOL depends on the interaction between the toric IOL and the posterior capsule, whereas misalignment of the toric IOL (defined in this study as the difference between the 3 months postoperatively measured axis of the toric IOL and the intended axis) depends on several factors additionally to rotational stability. Intraoperatively, misalignment may happen due to cyclotorsion of the eye in the lying position or due to peribulbar anaesthesia, and due to imprecision of the surgeon when positioning of the IOL relative to the intended meridian. Both these imprecisions can be dealt with by pre-operative marking of the eye in the sitting position and diligence by the surgeon. Postoperatively, the IOL may rotate because it is undersized for the capsule bag or due to the capsule shrinkage that takes place during fibrotic contraction of the bag in the postoperative period. Since most current IOLs are slightly oversized for the capsule bag, the former is observed rarely and would be more likely in long eyes which tend to have a larger capsule bag diameter. However, capsule bag shrinkage is thought to induce rotation in IOLs with open-loop haptics due to the asymmetry of the haptic design. Typical IOL haptic designs that improve rotational stability are either plate haptic IOLs or special Z-haptic shaped open-loop haptics that attempt to counteract the rotational effect of compression of the shrinking bag. However, both of these IOL designs are thought to have downsides. The former seems to have an increased risk of posterior capsule opacification due to a less effective lens epithelial barrier effect of the optic edge and may also show more rotation in the bag immediately after surgery due to the shorter haptic overall length. The latter IOL design, such as the Z-haptic IOL, is cumbersome to implant and may be prone to damage during implantation.

Aim of this study was to evaluate the rotational stability of a novel single-piece hydrophobic acrylic toric IOL with a C-haptic design.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
21 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • corneal astigmatism between 1.0 D and 3.0 D

排除标准

  • irregular astigmatism
  • forme fruste Keratoconus
  • corneal scars
  • phakodonesis
  • pseudoexfoliation syndrome
  • traumatic cataract

结局指标

主要结局

Rotational stability of the investigated toric lens

时间窗: 3 months

Rotation of the investigated intraocular lens using a photograph technique (retroillumination images) immediately after surgery and 3 months later.

次要结局

  • reduction in post-operative higher order aberrations(3 months)

研究者

发起方
Prim. Prof. Dr. Oliver Findl, MBA
申办方类型
Other
责任方
Sponsor Investigator
主要研究者

Prim. Prof. Dr. Oliver Findl, MBA

Prim. Prof. Dr., MBA

Vienna Institute for Research in Ocular Surgery