The Clinical Evaluation of the Cinatra™ Corolimus-Eluting Coronary Stent in De Novo Lesions in Native Coronary Arteries

Not Applicable
Terminated
Conditions
Coronary Artery Disease
Interventions
Device: Cinatra™ Corolimus Eluting Coronary Stent System
Registration Number
NCT01025869
Lead Sponsor
Atrium Medical Corporation
Brief Summary

To investigate the safety and efficacy of the Cinatra™ Corolimus Drug Eluting Stent for the treatment of de novo lesions in native coronary arteries.

Detailed Description

This is a single-arm, multicentre pilot study designed to provide an indication of the effectiveness and safety of the Cinatra™ Corolimus Eluting Coronary Stent System. The primary endpoint to be evaluated in this study is late lumen loss (in-stent) at 6 months post-procedure as measured by QCA in the 30 participants undergoing angiography at this timepoint. Late lumen loss is defined as the difference between the post-index procedure minimal lumen diameter (MLD) and the follow-up MLD.

Recruitment & Eligibility

Status
TERMINATED
Sex
All
Target Recruitment
31
Inclusion Criteria
  1. Patient is ≥ 18 years old
  2. Patient is an acceptable candidate for percutaneous coronary intervention (PCI), stenting, and emergent coronary artery bypass graft (CABG) surgery
  3. Patient has clinical evidence of ischemic heart disease, stable or unstable angina, silent ischemia, and/or a positive functional study
  4. Female subjects of childbearing potential must have a negative pregnancy test within 7 days before the trial procedure
  5. Patient or subject's legal representative has been informed of the nature of the trial and agrees to its provisions and has provided written informed consent as approved by the Hospital Research Ethics Committee (HREC) of the respective investigational site
  6. Patient agrees to comply with specified follow-up evaluations and to return to the same investigational site where the procedure was performed

Angiographic:

  1. Patient has either a single target lesion, or two lesions (target and non-target) located in separate coronary arteries

  2. If a non-target lesion is treated, it must be treated first and only with commercially available PTCA balloons and/or stents. Post PCI of the non-target vessel, all of the following conditions must be met:

    • Residual diameter stenosis <10%
    • Absence of any angiographic complications
    • Absence of ischaemic symptoms
    • Absence of significant new arrhythmia or ECG monitoring changes suggestive of ischaemia
  3. Target lesion must be a de novo lesion in native coronary artery

  4. Target lesion must be ≤ 22 mm in length

  5. Target lesion must have a stenosis of ≥ 50% and < 100%

  6. Target vessel must have a reference vessel diameter (RVD) appropriate for treatment with a of 3.0mm or3.5 mm stent

  7. Target vessel must have a Thrombolysis in Myocardial Infarction (TIMI) flow ≥ 2

Exclusion Criteria
  1. Known hypersensitivity or contraindication to aspirin, heparin or bivalirudin, clopidogrel or ticlopidine, cobalt, chromium, stent coatings (i.e. fatty acids, glycerides, and alpha tocopherol), or a sensitivity to contrast media, which cannot be adequately pre-medicated

  2. History of an allergic reaction or significant sensitivity to drugs such as , zotarolimus, rapamycin, tacrolimus, everolimus, or any other analogue or derivative

  3. Platelet count < 100,000 cells/mm³ or > 700,000 cells/mm³, or a white blood cell (WBC) count < 3,000 cells/mm³ within 7 days prior to index procedure

  4. Serum creatinine level 170 micromol/L within 7 days prior to index procedure

  5. Evidence of an acute myocardial infarction (MI) within 72 hours of the intended trial procedure (defined as: Q wave myocardial infarction (QWMI) or non-Q wave myocardial infarction (NQWMI) having CK enzymes > 2X the laboratory upper limit of normal with the presence of an elevated CK-MB (any amount above the laboratory upper limit of normal)

  6. Previous PCI of the target vessel within 9 months prior to the procedure

  7. Any planned additional PCI procedure within 30 days post-index procedure and/or planned PCI of the target vessel within 12 months post-procedure

  8. During the index procedure, the target and/or non-target lesion(s) requires treatment with a device other than PTCA prior to stent placement (including, but not limited to, cutting balloon, atherectomy, thrombectomy, etc.)

  9. Left ventricular ejection fraction (LVEF) < 30% if evaluated, or clinical evidence of significant congestive heart failure (NYHA Class III or IV) within the prior 30 days

  10. History of a stroke or transient ischemic attack (TIA) within the prior 6 months

  11. Active peptic ulcer or upper gastrointestinal (GI) bleeding within the prior 6 months

  12. History of bleeding diathesis or coagulopathy or will refuse blood transfusions

  13. Concurrent medical condition with a life expectancy of less than 12 months

  14. Any previous treatment of the target vessel(s) for restenosis, including brachytherapy

  15. Any condition which, in the Investigator's opinion, may interfere with the subject's optimal participation in the study 17. Currently participating in an investigational drug or another device trial that has not completed the primary endpoint or that clinically interferes with the current trial endpoints; or requires coronary angiography, IVUS or other coronary artery imaging procedures

Angiographic:

  1. Target lesion is located in native vessel distal to anastomosis with a saphenous vein graft or a left/right internal mammary artery (LIMA/RIMA) bypass with more than 40% diameter stenosis anywhere within the graft

  2. Previous stenting in the target vessel.

  3. The target vessel has other lesions with greater than 40% diameter stenosis based on visual estimate or on-line QCA

  4. The target vessel has evidence of thrombus

  5. The target vessel is excessively tortuous (two bends > 90º to reach the target lesion)

  6. The target lesion has any of the following characteristics:

    • Lesion location is aorto-ostial, an unprotected left main lesion, or within 5 mm of the origin of the left anterior descending (LAD) or left circumflex (LCX)
    • Involves a side branch > 2.0 mm in diameter
    • Is at or distal to a > 45º bend in the vessel
    • Is severely calcified
  7. Unprotected left main coronary artery disease (an obstruction greater than 50% in the left main coronary artery)

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Arm && Interventions
GroupInterventionDescription
Stent SystemCinatra™ Corolimus Eluting Coronary Stent SystemCinatra™ Corolimus Eluting Coronary Stent System
Primary Outcome Measures
NameTimeMethod
In-stent late lumen loss (LLL) as measured by quantitative coronary angiography (QCA).6 months post treatment
Secondary Outcome Measures
NameTimeMethod
Target lesion revascularization1, 6 and 18 month and annually to 5 years
Target vessel revascularization1 month and at all follow up to 5 years
Stent thrombosisAll follow ups
Neointimal Hyperplasia6 and 18 months
MACE (Major Adverse Cardiac Event)1 month, 6 month, 18 month and annually to 5 years
In-segment late lumen loss (LLL) as measured by quantitative coronary angiography6 and 18 months
In-stent late lumen loss (LLL) as measured by quantitative coronary angiography18 months (optional)
Minimal Lumen Diameter (MLD), in-stent and in-segment6 and 18 months
Rates of incomplete stent apposition6 and 18 months
Device success defined as achievement of a final residual diameter stenosis of < 30% measured by QCA, using the study device only.procedure
Lesion treatment success is defined as <30% residual stenosis measured by QCA by any treatmentProcedure
Procedure success defined as lesion success without the occurrence of MACE during the hospital staydischarge
Binary restenosis6 and 18 months

Trial Locations

Locations (4)

Auckland City Hospital

🇳🇿

Auckland, New Zealand

North Shore Hospital

🇳🇿

Auckland, New Zealand

Christchurch Hospital

🇳🇿

Christchurch, New Zealand

Mercy Angiography

🇳🇿

Auckland, New Zealand