An Open Label, Interventional, Clinical Performance and Validation Study to Demonstrate the Clinical Accuracy of of the V-Sensor in the Healthy Adult Population; Single Site in Switzerland

Léman Micro Devices SA0 sites95 target enrollmentApril 2018

Overview

Phase
N/A
Intervention
Not specified
Conditions
Clinical Accuracy of Vital Sign Pressure Sensor
Sponsor
Léman Micro Devices SA
Enrollment
95
Primary Endpoint
Pulse Rate accuracy to the reference device within the ranges specified by the ISO 80601-2-61
Last Updated
8 years ago

Overview

Brief Summary

This trial seeks to demonstrate clinical accuracy and safety of the integrated V-Sensor when used by the general, healthy adult population interested in measuring their vital signs outside of the clinical setting.

Detailed Description

The five vital signs are blood pressure, pulse, respiration, SpO2, and temperature. The user tests two devices mounted on two smartphones. * Temperature: Each user will use the V-sensor mounted on a smartphone to obtain his/her temperature interfaced with a temperature specific 'app'. Before and after the the temperature is obtained using the V-Sensor, the temperature will be obtained using an equivalent thermometer. * Blood Pressure: Each user will test his/her blood pressure using the two sensors mounted on the two phones interfaced with a blood pressure specific 'app'. Before and after each measurement, a reference measurement using a mercury sphygmomanometer will be taken. * Pulse, Respiration, SpO2: Each user will test his/her pulse, respiration, and SpO2 using 2 V-sensors mounted on two separate smartphones interfaced with a Pulse, Respiration, and SpO2 'app'. Before and after each measurement, a reference measurement using equivalent reference devices will be taken.

Registry
clinicaltrials.gov
Start Date
April 2018
End Date
September 2018
Last Updated
8 years ago
Study Type
Interventional
Study Design
Single Group
Sex
All

Investigators

Sponsor
Léman Micro Devices SA
Responsible Party
Sponsor

Eligibility Criteria

Inclusion Criteria

  • Informed Consent
  • 18 and 80 years old
  • no history or risk for developing lymphedema
  • no arteriovenous fistula in the right arm
  • no right limb amputation up to and including the right index finger
  • ability to obtain 3 readings from the device during the screening/training session.

Exclusion Criteria

  • history of arrhythmia
  • arteriovenous fistula in the right arm
  • lymphedema or high risk for developing lymphedema in the right arm
  • inability to follow the procedures of the study
  • inability to hold the study device as instructed
  • inability to obtain 3 readings from the study device during the screening session
  • previous enrolment into the current study
  • enrolment of the Investigator or Sponsor, his/her family members, employees and other dependent persons.

Outcomes

Primary Outcomes

Pulse Rate accuracy to the reference device within the ranges specified by the ISO 80601-2-61

Time Frame: 4 months

Pulse rate accuracy shall be stated as the root-mean-square (rms) difference between paired pulse rate data recorded with the pulse oximeter equipment and with the Nellcor N-PA reference device.

Respiration rate accuracy

Time Frame: 4 months

Respiration rate accuracy shall be stated as the root-mean-square (rms) difference between paired respiratory rate data with the Passport Respiration Rate Sensor PS-2133 reference method; target range +/- 10% of participant's respiratory rate

Temperature accuracy to the reference device with the ranges specified by the ISO 80601-2-56 & ATSM 1965 Standard.

Time Frame: 4 months

Temperature accuracy shall be stated as the root-mean-square (rms) difference between paired temperatures data recorded with the Exergen Temporal Scanner thermometer. LMD's target is a clinical bias \<= 3.0°C and a clinical repeatability \<= 3.0°C.

Blood Pressure accuracy to the reference device within the ranges specified by the ISO standard 81060-2

Time Frame: 4 months

For systolic and diastolic blood pressures, the mean error of determination of the individual paired determinations of the sphygmomanometer-under-test and the reference sphygmomanometer (Standby Baumanometer) for all participants shall not be greater than 8.0mmHg.

Blood Oxygen accuracy to the reference device within the ranges specified by the ISO 80601-2-61.

Time Frame: 4 months

The SpO2 accuracy shall be stated as a root-mean-square (rms) difference of less than or equal to 4.0% SpO2 over the range of 70% to 100% SaO2

Secondary Outcomes

  • No serious adverse device effects are reported.(4 months)