Non-Contact Air Tonometry (Topcon CT-1®) Versus Goldmann
- Conditions
- Glaucoma
- Registration Number
- NCT03684174
- Lead Sponsor
- University of Zurich
- Brief Summary
In this prospective device comparison study with 58 healthy eyes, the Topcon CT-1® showed statistically significant elevated intraocular pressure measurements compared to the gold standard Goldmann application tonometry.
- Detailed Description
Introduction:
Intraocular pressure (IOP) measurement is crucial in glaucoma diagnosis and treatment. We compared a non-contact air tonometer (Topcon CT-1®) with the gold standard Goldmann Applikation tonometer (GAT).
Methods:
In this prospective, mono centric device comparison study, only healthy eyes were enrolled. IOP was measured three times with the Topcon CT-1® followed by three times GAT. Mean IOP was calculated and compared with a paired t-test, a linear regression and a Bland-Altman analysis.
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 58
Healthy volunteers who agreed to be included in the study and were able to comply with the measurement procedures
Exclusion criteria were cor-neal abnormalities (such as oedema, scars or dystrophies, astigmatism,hyperopia or myopia > 3 dioptres), a history of ocular surgery or recent(<47 weeks) ocular disorder.
Not provided
Study & Design
- Study Type
- OBSERVATIONAL
- Study Design
- Not specified
- Primary Outcome Measures
Name Time Method Comparison of GAT with air tonometry Topcon CT1 7 months Difference in mmHg between these two devices
- Secondary Outcome Measures
Name Time Method
Trial Locations
- Locations (1)
Augenpraxis Centerpassage
🇨🇭Wettingen, AG, Switzerland
Augenpraxis Centerpassage🇨🇭Wettingen, AG, Switzerland