Non-Contact Air Tonometry (Topcon CT-1®) Versus Goldmann

Completed
Conditions
Glaucoma
Registration Number
NCT03684174
Lead Sponsor
University of Zurich
Brief Summary

In this prospective device comparison study with 58 healthy eyes, the Topcon CT-1® showed statistically significant elevated intraocular pressure measurements compared to the gold standard Goldmann application tonometry.

Detailed Description

Introduction:

Intraocular pressure (IOP) measurement is crucial in glaucoma diagnosis and treatment. We compared a non-contact air tonometer (Topcon CT-1®) with the gold standard Goldmann Applikation tonometer (GAT).

Methods:

In this prospective, mono centric device comparison study, only healthy eyes were enrolled. IOP was measured three times with the Topcon CT-1® followed by three times GAT. Mean IOP was calculated and compared with a paired t-test, a linear regression and a Bland-Altman analysis.

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
58
Inclusion Criteria

Healthy volunteers who agreed to be included in the study and were able to comply with the measurement procedures

Exclusion criteria were cor-neal abnormalities (such as oedema, scars or dystrophies, astigmatism,hyperopia or myopia > 3 dioptres), a history of ocular surgery or recent(<47 weeks) ocular disorder.

Exclusion Criteria

Not provided

Study & Design

Study Type
OBSERVATIONAL
Study Design
Not specified
Primary Outcome Measures
NameTimeMethod
Comparison of GAT with air tonometry Topcon CT17 months

Difference in mmHg between these two devices

Secondary Outcome Measures
NameTimeMethod

Trial Locations

Locations (1)

Augenpraxis Centerpassage

🇨🇭

Wettingen, AG, Switzerland

Augenpraxis Centerpassage
🇨🇭Wettingen, AG, Switzerland