Patient-ventilator Synchronisation Study in Non Invasive Ventilation for Intensive Care Unit Patients: Comparison Between Manual and Automated Ventilator Settings.
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 16
- 试验地点
- 2
- 主要终点
- Asynchrony index
研究概览
简要总结
This cross-over study will compare the asynchrony index between standard manual ventilator settings, optimized manual ventilator settings, and automated ventilator setting in intensive care patients ventilated in non-invasive ventilation with a high asynchrony index. The hypothesis is that both manual optimized ventilator settings and automated ventilator settings are associated with a lower patient-ventilator asynchrony index as compared to manual standard ventilator settings.
A randomized cross-over design method will be used. Patient requiring NIV with an asynchrony index over 35% will be included. An esophageal catheter with a balloon will be inserted to monitor esophageal pressure. Patients will be ventilated during 3 periods of 30 min, with 10 minutes of washout in between. Recordings of airway pressure, airway flow, and esophageal pressure will be analyzed by two investigators blinded of the trigger settings.
The primary outcome will be the asynchrony index. The secondary outcome will be the ineffective inspiratory effort index, autotrigering index, double triggering index, inspiratory trigger delay, cycling delay, total time spent in asynchrony, patient comfort, and blood gas results.
详细描述
Non-invasive ventilation (NIV) is used in 35% of patient admitted in intensive care unit (ICU) with a failure rate of 10 to 70% depending on the indication and clinician experience. Patient-ventilator asynchrony is a frequent cause of NIV failure. Therefore, optimizing patient-ventilator synchronization is important for its comfort, tolerance, and efficacy. An optimal patient-ventilation is achieved when the mechanical breath provided by the ventilator match the patient inspiratory effort. The ratio between the number of asynchronies divided by the number of patient inspiratory effort define the asynchrony index (AI). AI over 10% is considered as severe and occurs in 30 to 43% of patients ventilated in NIV. Patient ventilator asynchronies occurs because ventilator settings of inspiratory and expiratory triggers remain constant in patient with variable respiratory drive, and unintentionnals leaks that are difficult to control in NIV. Thus using an automatic adjustment of inspiratory and expiratory triggers setting according to patient effort and unintentional leaks may decrease the number of patient-ventilator asynchronies. This cross-over study will compare the asynchrony index between standard manual ventilator settings, optimized manual ventilator settings, and automated ventilator setting in intensive care patients ventilated in non-invasive ventilation with a high asynchrony index. The hypothesis is that both manual optimized ventilator settings and automated ventilator settings are associated with a lower patient-ventilator asynchrony index as compared to manual standard ventilator settings.
A randomized cross-over design method will be used. Patient requiring NIV with an asynchrony index over 30% will be included. An esophageal catheter with a balloon will be inserted to monitor esophageal pressure. Patients will be ventilated during 3 periods of 30 min, with 10 minutes of washout in between. Recordings of airway pressure, airway flow, and esophageal pressure will be analyzed by two investigators blinded of the trigger settings.
The primary outcome will be the asynchrony index. The secondary outcome will be the ineffective inspiratory effort index, autotrigering index, double triggering index, inspiratory trigger delay, cycling delay, total time spent in asynchrony, patient comfort, and blood gas results.
The sample size was calculated from the total asynchrony index (primary outcome). Patients with an asynchrony index over 30% in using manual standard ventilator settings will be included. Considering an asynchrony index of 30 ± 15 % in manual standard ventilator settings with a clinically significant objective to reduce the asynchrony index to 15% in manual optimized ventilator settings and automated ventilator settings, a sample size of 30 patients is required with a risk at 0.05 and a power at 80%. Therefore, 35 patients are planned.
研究设计
- 研究类型
- Interventional
- 分配方式
- Randomized
- 干预模型
- Crossover
- 主要目的
- Treatment
- 盲法
- Single (Participant)
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Patient aged over 18 years old
- •Covered by social insurance
- •Consent for study signed by patient or next-of-kin
- •NIV session indicated for at least 2 hours
- •Asynchrony index ≥ 30% with standard manual settings
排除标准
- •Patient requiring continuous NIV
- •Contra-indication to esophageal catheter insertion: gastric ulcer, esophageal varices, pharyngeal or laryngeal tumor.
- •Patient with withholding decision about intubation
- •Moribund patient
- •Patient included in another interventional study in the last 30 days
- •Patient that does not speak French
- •Pregnant women
结局指标
主要结局
Asynchrony index
时间窗: Continuous measurement over 30min
Ratio between the total number of asynchronies divided by the number of patient inspiratory effort
次要结局
- Double triggering index(Continuous measurement over 30min)
- Blood PaO2 results(After each period at 30 min, 1 h and 1 h 30 min)
- Blood PaCO2 results(After each period at 30 min, 1 h and 1 h 30 min)
- Blood pH results(After each period at 30 min, 1 h and 1 h 30 min)
- Total time spent in asynchrony(Continuous measurement over 30min)
- Lineffective inspiratory effort index(Continuous measurement over 30min)
- Autotrigering index(Continuous measurement over 30min)
- Inspiratory trigger delay(Continuous measurement over 30min)
- Cycling delay(Continuous measurement over 30min)
- Patient comfort(1 day (Single measurement))