Verification of Pulse Oximetry in Neonatal Population

Children's Hospitals and Clinics of Minnesota1 个研究点 分布在 1 个国家目标入组 25 人开始时间: 2008年6月最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
25
试验地点
1

研究概览

简要总结

The purpose of this his post-market, observational study titled Verification of Pulse Oximetry in the Neonatal Population is to gather data to further support the accuracy of Nonin's pulse oximeters when used in the neonate population. Neonates who are currently undergoing arterial blood gas (ABG) sampling, as part of their medical care will be recruited to participate in this study for collecting opportunity samples. Nonin's PureLight® noninvasive sensor will be applied to a foot of the neonate; an adhesive wrap secures the sensor to the neonate. The sensor immediately begins obtaining SpO2 and pulse rate measurements which are sent to the oximeter module for display on the Electronic Data Collection System (EDCS). At the time the prescribed ABG is drawn, the EDCS must have a marker captured to identify the end of the ABG for SpO2 comparison. The SpO2 accuracy of pulse oximeter will be measured by comparing SpO2 readings of the pulse oximeter to values of SaO2 (opportunity samples) determined by a CO-oximeter.

研究设计

研究类型
Observational
观察模型
Case Control
时间视角
Prospective

入排标准

年龄范围
1 Day 至 30 Days(Child)
性别
All
接受健康志愿者

入选标准

  • Gestational age: 25 weeks to full term
  • Weight range: No minimum or maximum range
  • Age: up to 1 month (30 DAYS)
  • Gender: Both
  • ABGs as part of their medical care, to be analyzed by co-oximeter
  • Consent signed by at least one legal guardian

排除标准

  • Presence of any skin irritation or breakdown on the foot
  • Foot impediments which would preclude proper placement of the sensor
  • Clinical condition precluding the use of adhesive materials
  • Cessation of ABGs during study enrollment

研究者

研究点 (1)

Loading locations...