MedPath

SIRONA 2 Trial Heart Failure NYHA Class III

Not Applicable
Active, not recruiting
Conditions
Heart Failure NYHA Class III
Registration Number
NCT04012944
Lead Sponsor
Endotronix, Inc.
Brief Summary

This is a prospective, multi-center, open-label, single-arm CE-Mark trial to assess device safety and efficacy of the Cordella PA Sensor System in up to 75 New York Heart Association (NYHA) Class III Heart Failure patients who will receive the Cordella PA Sensor Implant.

Detailed Description

The study objectives are:

1. To establish that the Cordella PA Sensor can be safely delivered, deployed, and remain stable within the target pulmonary artery (PA) segment through 30 days post-implant.

2. To compare Cordella PA Sensor System pressure measurements with fluid-filled catheter PA pressure measurements obtained by standard right heart catheterization (RHC) at 90 days post implant.

Recruitment & Eligibility

Status
ACTIVE_NOT_RECRUITING
Sex
All
Target Recruitment
81
Inclusion Criteria
  1. Subject has given written informed consent

  2. Male or female, at least 18 years of age

  3. Diagnosis of HF ≥ 6 months with either preserved or reduced left ventricular ejection fraction (LVEF) and NYHA Class III HF at the time of Screening

  4. HF related hospitalization, HF treatment in a hospital day-care setting, or unplanned outpatient clinic HF visit within 12 months prior to consent and/or increase of N-terminal pro-Brain Natriuretic Peptide (NT-proBNP) (or Brain Natriuretic Peptide (BNP)) at time of Screening defined as:

    1. Subjects with LVEF ≤ 40%: NT-proBNP ≥ 1000 pg/mL (or BNP ≥ 250 pg/mL).
    2. Subjects with LVEF > 40%: NT-proBNP ≥ 700 pg/mL (or BNP ≥ 175 pg/mL). Thresholds for NT-proBNP and BNP (for both LVEF ≤ 40% and LVEF > 40%) will be corrected for body mass index (BMI) using a 4% reduction per BMI unit over 25 kg/m2
  5. Subjects must be receiving appropriate medical management of HF for 3 months prior to Screening and clinically stable for at least 1 month prior to study entry

  6. Subjects who are physically able to hold the myCordella™ Patient Reader unit (approximate weight 600g) against the ventral thoracic surface for up to 2 minutes per day while in a seated position, as well as dock and undock the myCordella™ Patient Reader

  7. Subjects with sufficient eyesight, hearing, and mental capacity to respond to the myCordella™ Patient Reader's audio/visual cues and operate the myCordella™ Patient Reader

  8. Subject has sufficient Cellular and/or Wi- Fi Internet coverage at home

  9. Subject agrees to return to the treating Investigator for all scheduled follow up visits and can return to the hospital for follow up

Exclusion Criteria
  1. Subjects with primary pulmonary hypertension
  2. Subjects with an active infection at the Cordella PA Sensor Implant Visit
  3. Subjects with history of pulmonary embolism or deep vein thrombosis
  4. Subjects who have had a major cardiovascular event (e.g., myocardial infarction, stroke) within 2 months of the Screening Visit
  5. Subjects whereby RHC is contraindicated
  6. Implanted with a Cardiac Rhythm Management (CRM) Device (Internal Cardiac Defibrillator (ICD) or pacemaker) or Cardiac Resynchronization Therapy (CRT)-Pacemaker (CRT-P) or CRT-Defibrillator (CRT-D) less than 90 days prior to screening visit
  7. Any major surgery within 30 days of the Sensor Implant Visit.
  8. Subjects with a Glomerular Filtration Rate (GFR) <25 ml/min or who are on chronic renal dialysis
  9. Specific liver enzymes [Aspartate Aminotransferase (AST) (SGOT), and Alanine Aminotransferase (ALT) (SGPT) >3 times the upper limit of normal
  10. Received or are likely to receive an advanced therapy (e.g., mechanical circulatory support or lung or heart transplant) in the next 12 months
  11. Subjects with congenital heart disease or mechanical/tissue right heart valve(s)
  12. Subjects with known coagulation disorders
  13. Subjects with a hypersensitivity or allergy to platelet aggregation inhibitors including aspirin, clopidogrel, prasugrel, and ticagrelor; or patients unable to take dual antiplatelet or anticoagulants for one- month post implant
  14. Known history of life threatening allergy to contrast dye
  15. Subjects who are pregnant or breastfeeding
  16. Subjects who are unwilling or deemed by the Investigator to be unwilling to comply with the study protocol, or subjects with a history of non-compliance
  17. Severe illness, other than heart disease, which would limit survival to <1 year
  18. Subjects whose clinical condition, in the opinion of the Investigator, makes them an unsuitable candidate for the study
  19. Subjects enrolled in another investigational trial with an active Treatment Arm
  20. Subject who is in custody by order of an authority or a court of law
  21. Unrepaired severe valvular disease
  22. Subjects with an inferior vena cava (IVC) filter

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Primary Outcome Measures
NameTimeMethod
Safety: Freedom from Adverse Events30 days

To evaluate the primary safety endpoint: Freedom from AEs associated with use of the CorPASS through 30 days post Cordella PA Sensor implant, the number of subjects with and without any Adverse Device Events will be summarized.

Accuracy of Cordella PA Sensor System pressure measurements, compared to standard-of-care fluid-filled catheter PA pressure measurements obtained by standard Right Heart Catheterization (RHC)90 days

System accuracy will be compared to the accuracy of standard-of-care commercial products that use fluid-filled invasive catheters to measure PA pressure at 90 days post implant visit.

Secondary Outcome Measures
NameTimeMethod
Device/system-related complications30 days and 2 years

The rate of device/system-related complications will be assessed throughout the study

Accuracy of Cordella PA Sensor pressure measurements12 months

Accuracy of Cordella PA Sensor System pressure measurements, compared to standard-of-care fluid-filled catheter PA pressure measurements obtained by standard RHC

HF related hospitalizations, treatments in a hospital day-care setting, or urgent outpatient clinic visits.90 days and 2 years

Frequency of HF hospitalizations, HF treatments in a hospital day-care setting, or urgent outpatient clinic HF visits.

Quality of Life measured by Kansas City Cardiomyopathy Questionnaire (KCCQ)90 days and 2 years

The Quality of Life KCCQ questionnaire is a measure of health-related quality of life. It is a 23-item, self-administered instrument that quantifies physical function, symptoms (frequency, severity and recent change), social function, self-efficacy and knowledge, and quality of life related to Heart Failure. Out of 23-items, 10 scores will be calculated: Physical Limitation, Symptom Stability, Symptom Frequency, Symptom Burden, Total Symptom Score, Self-Efficacy, Quality of Life, Social Limitation, Overall Summary Score, Clinical Summary Score .Presentation will be done by means of descriptive statistics including the absolute value and change from baseline by visit.

Functional Status Change as measured by 6-Minute Walk Test90 days and 2 years

Subjects complete a six (6) minute walk test at the study visits described in the study protocol. Total distance walked (meters) will be collected and analyzed by means of descriptive statistics including the absolute value and change from baseline by visit.

Pressure sensor failure rate30 days and 2 years

Pressure sensor failure rate throughout the study

Heart Failure related medication changes90 days and 2 years

Heart failure related medication change will be collected and summarized by visit presenting the percentage of subjects with and without any change in Heart Failure related medication

Device Success90 days

Percentage of device success as documented by ability of the System to successfully transmit collected PAP data to a secure database

Change in Pulmonary Artery (PA) pressure90 days and 2 years

Change in PA pressure:

1. Pre- and post-implant

2. Before and after 6-Minute Walk Test

3. Sitting vs supine PA pressure measurements

Adherence to regular myCordella™ Peripherals measurements measured as a percentage of subjects who adhere to regular myCordella™ Peripherals measurements90 days and 2 years

Subjects will be instructed to perform daily measurements throughout the study. Subject compliance will be defined as at least 5 days of data collection/transmissions per week.

Functional Status Change as measured by New York Heart Association (NYHA) functional classification90 days and 2 years

The NYHA functional classification categorizes the extent of heart failure by placing subjects in one of four (I, II, III, IV) categories based on how much they are limited during physical activity and symptoms of shortness of breath and/or angina. Shifts in NYHA functional classification from baseline over all post-baseline visits will be presented.

Accuracy using the Bland Altman Method90 days

Accuracy of Cordella PA Sensor System pressure measurements, compared to standard-of-care fluid-filled catheter PA pressure measurements obtained by standard RHC using the Bland Altman method.

Frequency of Adverse Events30 days and 2 years

Frequency of adverse events throughout the study

Trial Locations

Locations (7)

Cardivascular Center OLV Aalst

🇧🇪

Aalst, Belgium

Ziekenhuis- Oost Limburg

🇧🇪

Genk, Belgium

Uniklinik Köln, Klinik III für Innere Medizin Herzzentrum der Universität zu Köln

🇩🇪

Cologne, Germany

Universitätsklinikum Giessen und Marburg GmbH, Standort Giessen Medizinische Klinik I, Abteilung Kardiologie und Angiologie

🇩🇪

Gießen, Germany

Klinik für Kardiologie und Angiologie/ Studienambulanz Medizinische Hochschule Hannover (MHH)

🇩🇪

Hannover, Germany

Mater Misericordiae University Hospital

🇮🇪

Dublin, Ireland

University Hospital Galway

🇮🇪

Dublin, Ireland

Cardivascular Center OLV Aalst
🇧🇪Aalst, Belgium

MedPath

Empowering clinical research with data-driven insights and AI-powered tools.