Skip to main content
Clinical Trials/NCT05791032
NCT05791032
Completed
Not Applicable

Sub Chronic Evaluation for ATP With an Extravascular Placed ICD Lead (STEP ICD) Study

AtaCor Medical, Inc.1 site in 1 country21 target enrollmentSeptember 5, 2023

Overview

Phase
Not Applicable
Intervention
Not specified
Conditions
Ventricular Arrythmia
Sponsor
AtaCor Medical, Inc.
Enrollment
21
Locations
1
Primary Endpoint
Induced VF Conversion Success
Status
Completed
Last Updated
2 years ago

Overview

Brief Summary

STEP ICD is a premarket, exploratory, early feasibility, interventional study designed to evaluate the preliminary safety and performance of the Investigational Devices. The study is intended to inform the final device design which will be further evaluated in traditional feasibility and /or pivotal clinical investigations.

The primary safety objective is to characterize safety of the EV-ICD Lead through 3 months post-implant. The primary performance objective is to characterize sensing and conversion of induced VF with the EV-ICD Lead up to 3 months post-implant.

Detailed Description

The primary safety endpoint is incidence of Adverse Device Effects (ADEs). The primary performance endpoint is induced VF conversion success up to 3 months post-implant. Secondary endpoints include defibrillation metrics (i.e., lowest Conversion Energy (J), ability to automatically sense and detect ventricular arrhythmias, detection time and time-to-therapy) and pacing metrics (ventricular capture threshold (V), pacing impedance (ohms), R-wave amplitude (mV), and extracardiac pacing sensation level). Up to 30 eligible Subjects undergoing a pectoral transvenous ICD procedure (de novo or replacement) will receive an EV-ICD Lead connected to a commercially available ICD in a left mid-axillary or left pectoral pocket. The EV-ICD System will be tested in the intended implant location to determine the lowest defibrillation energy. Subjects will remain in the hospital overnight (at a minimum), prior to discharge. Except during testing, the EV-ICD System will remain programmed to monitor only (therapy OFF) to detect and store ventricular arrhythmias over the follow-up period. A concomitantly implanted transvenous ICD (TV-ICD) system will provide any necessary ICD therapy. Non-therapy EV-ICD programming parameters will match the TV-ICD to the extent possible to facilitate recording of matched stored electrograms. Follow-up visits are planned at pre-discharge, 2 weeks, 1 month, 2 months, and 3 months. VF conversion testing will occur prior to lead removal in order to test subchronic defibrillation effectiveness with a submaximal energy shock in standard and/or reverse polarity. At the Lead Removal Follow Up, the EV-ICD System will be explanted and the TV-ICD System left in place. A final follow-up will occur one month after EV-ICD Lead explant to assess any post-explant adverse events prior to study exit.

Registry
clinicaltrials.gov
Start Date
September 5, 2023
End Date
March 19, 2024
Last Updated
2 years ago
Study Type
Interventional
Study Design
Single Group
Sex
All

Investigators

Responsible Party
Sponsor

Eligibility Criteria

Inclusion Criteria

  • At least 18 years old
  • Indicated for de novo or replacement ICD procedure

Exclusion Criteria

  • BMI ≥ 35 kg/m2
  • Subjects who require continual ventricular pacing
  • Presence or planned use of a subcutaneous ICD lead, subcutaneous coils/arrays, epicardial patches or epicardial pace/sense leads within the study period
  • Planned MRI within the study period
  • Subjects on anticoagulation or antiplatelet therapy that cannot be temporarily discontinued for the procedure
  • Circumstances that may prevent data collection or follow-up
  • Participation in any concurrent clinical study without prior written approval from the Sponsor
  • Inability or unwillingness to provide informed consent to participate in the Study
  • Known prior history for any of the following:
  • NYHA IV functional class in past 90 days

Outcomes

Primary Outcomes

Induced VF Conversion Success

Time Frame: Up to 3 months post-implant

Induced VF conversion success up to 3 months post-implant

Incidence of ADEs

Time Frame: Up to 3 months post-implant

Incidence of Adverse Device Effects (ADEs)

Study Sites (1)

Loading locations...

Similar Trials