A Post-Market Surveillance Clinical Registry of the TREOVANCE® Stent-Graft for Patients With Infrarenal Abdominal Aortic Aneurysms

Bolton Medical32 sites in 17 countries202 target enrollmentFebruary 5, 2014

Overview

Phase
Not Applicable
Intervention
Not specified
Conditions
Abdominal Aortic Aneurysm
Sponsor
Bolton Medical
Enrollment
202
Locations
32
Primary Endpoint
Number of subjects reporting Major Adverse Events
Status
Completed
Last Updated
5 years ago

Overview

Brief Summary

Patients with abdominal aortic aneurysms (AAA) suitable for endovascular aortic repair (EVAR) with Treovance were eligible to participate. Main inclusion criteria were: age 18-85 years; infrarenal AAA without significant infrarenal or distal iliac landing neck calcification or thrombus formation; infrarenal or distal iliac landing neck size requirements specified in the instructions for use. Main exclusion criteria: dissection/ruptured aneurysm or prior AAA endovascular or surgical repair. The primary endpoints were standard EVAR criteria.

Detailed Description

The goal of the registry was to collect clinical data of TREOVANCE® device in subjects with infrarenal aortic aneurysms, specifically to evaluate if the diseased pathology is treated with an acceptable technical success rate, an acceptable complication rate and that the device performs as expected. The following endovascular measures/parameters were assessed for preliminary performance at the follow-up intervals: delivery/deployment, stent-graft migration, stent-graft patency, stent-graft integrity, endoleak, aneurysm sac size changes, limb ischemia, and vascular access complications. This registry was a prospective, multi-center, post-market non-randomized study. The goal was to gather safety and performance data on the device. Subjects diagnosed with infrarenal aortic aneurysms enrolled into the registry were treated with the TREOVANCE® Stent-Graft with Navitel® Delivery System. Pre-procedure baseline data was gathered as well as post-procedure assessments prior to hospital discharge and 1, 6, and 12 months post-implantation.

Registry
clinicaltrials.gov
Start Date
February 5, 2014
End Date
April 11, 2017
Last Updated
5 years ago
Study Type
Observational
Sex
All

Investigators

Sponsor
Bolton Medical
Responsible Party
Sponsor

Eligibility Criteria

Inclusion Criteria

  • Subject must be between the ages of 18 and
  • Subject must be diagnosed with an infrarenal abdominal aortic aneurysm (AAA), with or without iliac artery involvement, by CT with contrast performed within 3 months of planned implant procedure.
  • Subject must have an infrarenal AAA that i. is \> 4.5 cm in diameter for females and \> 5 cm for males. ii. has increased in diameter by 0.5 cm in the last 6 months
  • Subject must have i. infrarenal landing neck length of 10 mm or greater and an angle of less than 60 degrees relative to the long axis of the aneurysm (centerline at lowest renal to centerline at bifurcation) and a suprarenal neck angle of less than 45 degrees relative to the infrarenal neck axis and an inside diameter of 17 mm - 32 mm, or ii. infrarenal landing neck length of 15 mm or greater and an angle of between 60 and 75 degrees relative to the long axis of the aneurysm and a suprarenal neck angle of less than 45 degrees relative to the infrarenal neck axis and an inside diameter of 16 mm-30 mm
  • Subject's infrarenal landing neck must i. have no significant calcification or thrombus formation, and ii. meet the vessel size requirements specified in the instructions for use (IFU) for the corresponding devices
  • Subject must have lowest renal artery at least 9 cm from the aortic bifurcation
  • Subject must have a distal iliac landing neck with i. an inside diameter of 7 mm - 13 mm and a length of at least 10 mm, or ii. an inside diameter of \>13 mm - 20 mm and a length of at least 15 mm
  • Subject's distal iliac landing neck must i. have no significant calcification or thrombus formation, and ii. meet the vessel size requirements specified for the corresponding devices in the IFU
  • Subject must have a total treatment length of at least 13 cm
  • Subject must be willing and able to comply with 1-month, 6-month, and 12-month follow-up visits.

Exclusion Criteria

  • Subject is pregnant or lactating
  • Subject has a dissection or a ruptured aneurysm (as determined by treating physician).
  • Subject has a patent inferior mesenteric artery that cannot be sacrificed and an occluded or stenotic celiac and/or superior mesenteric artery
  • Implant procedure as planned does not allow for at least one patent hypogastric artery left intact, unless both are occluded on pre-op imaging
  • Subject has a lesion that cannot be crossed by a guide wire
  • Proximal neck cannot increase by more than 20% over 15 mm; i.e., no conical necks
  • Subject has severe untreated coronary artery disease and/or unstable angina, significant areas of myocardium at risk (based on coronary angiogram or radionuclide scans), left ventricular ejection fraction \< 20%, or recent diagnosis of CHF
  • Subject has had a stroke or MI within 6 months of the planned treatment date
  • Subject has chronic obstructive pulmonary disease requiring routine need for oxygen therapy outside the hospital setting (e.g., daily or nightly home use)
  • Subject has an active systemic infection or is suspected of having an active systemic infection (e.g., AIDS/HIV, sepsis)

Outcomes

Primary Outcomes

Number of subjects reporting Major Adverse Events

Time Frame: Through 12 months

Primary safety outcome will be assessed by measurement of mortality and major morbidity rates through reporting of the following events: * Death * Stroke * Myocardial Infarction * Renal Failure requiring renal replacement therapy * Respiratory Failure, defined as ventilator-dependent * Paraparesis / Paraplegia (excludes paraparesis) * Bowel ischemia * Treated aneurysm rupture

Secondary Outcomes

  • Rate of Vascular Access Complications(1 month, 6 months and 12 months)
  • Rate of Stent-Graft Migration greater than 10mm(6 months and 12 months)
  • Rate of Endoleaks(1 month, 6 months and 12 months)
  • Rate of Aneurysm Sac Size Changes(6 months and 12 months)
  • Rate of unintentional covering of renal arteries and / or hypogastric arteries(Procedure / Study Day 0)
  • Rate of Device Delivery/Deployment Success(Procedure / Study Day 0)
  • Rate of Stent-Graft Patency(1 month, 6 months and 12 months)
  • Rate of Stent-Graft Integrity(1 month, 6 months and 12 months)
  • Rate of Limb Ischemia(1 month, 6 months and 12 months)