A Prospective Single-Site Non-Interventional Study to Evaluate the Diagnostic Accuracy and Repeatability of Pupil Measurements Using a Prototype Binocular Optical Coherence Tomography System.

University College, London1 个研究点 分布在 1 个国家目标入组 100 人开始时间: 2017年5月18日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
100
试验地点
1
主要终点
Diagnostic accuracy of the binocular OCT

研究概览

简要总结

In current, clinical ophthalmology, a range of specialised testing allows comprehensive evaluation of ocular health. These tests have typically evolved over many years to ensure their clinical validity. For example, the assessment of visual acuity has traditionally been measured with Snellen letter charts from a distance of six metres (20 feet), leading to the phrase "20/20 vision". Despite this, the limitations of Snellen testing are well established and more sophisticated testing is now available (e.g., logMAR testing using ETDRS (Early Treatment Diabetic Retinopathy Study) charts). Many other diagnostic tests have undergone similar cycles of refinement, often over extended time periods. Therefore, it should be incumbent on any new device to undergo detailed evaluation of its validity (how its measurements agree with other testing) and its repeatability (the variability when a further measurement is obtained in short time period, by the same operator and under the same conditions).

Binocular OCT extends the application of OCT devices beyond that of simple, cross- sectional imaging to a diverse array of diagnostic tests. The binocular design also removes the need for additional personnel to perform testing (i.e., the device can be self-operated in an automated manner), and allows for novel testing to be performed that is not possible with monocular imaging. In particular, binocular OCT devices have the potential to perform automated, quantitative pupillary measurements - an entirely novel application for this imaging modality.

This study will assess the validity and repeatability of pupil measurements using binocular OCT.

详细描述

Binocular OCT has the potential to allow for objective, quantitative measurement of pupillary reflexes, that can be performed in a fast, safe, and automated manner. Importantly, binocular OCT pupillometry can be performed alongside binocular retinal OCT imaging in a single testing session, in an automated manner (imaging of the retina using OCT is the commonest ophthalmic imaging procedure worldwide - in 2011, it is estimated that more than 20 million retinal OCT images were obtained, more than the sum of all other ophthalmic imaging procedures combined).

Assessment of pupillary reflexes is an essential component of a comprehensive ocular examination. In patients with visual loss, for example, detection of an RAPD often indicates the presence of serious ophthalmic diseases, such as optic neuropathies or severe retinal disease. As such, RAPD testing is essential as a screening test in the assessment of patients with less serious conditions such as cataract.

RAPD testing is commonly performed using the swinging flashlight method (SFM). A positive SFM test suggests the presence of asymmetric damage involving the afferent limb of the pupillary light reflex pathway. The SFM is a qualitative test, and relies upon visualising no pupillary constriction or immediate or delayed pupillary dilation; if either of these responses is observed, an RAPD is believed to be present. The SFM requires a trained examiner and many potential sources of error may confound the results, including anisocoria, off-axis pupil illumination, and unequal retinal bleaching. Pupil abnormalities may be subtle and easily missed using the SFM.

Automated pupilometers allow examiners to precisely and objectively quantify the pupillary response to light by measuring parameters such as minimum and maximum pupil diameter, the amplitude of constriction, latency of constriction, velocity of constriction, and duration of maximum constriction. Dedicated pupillometry devices are commercially available and have been shown to be highly sensitive and specific for RAPD detection e.g. RAPDx; Konan Medical USA, Inc., Irvine, CA. These devices are also capable of detecting the presence of more subtle RAPD, typically not found using the SFM, in diseases such as glaucoma.

Unfortunately, automated pupilometers such as the Konan RAPDx system are not widely used in clinical practice. This is largely because they are expensive devices, limited to a single purpose. A binocular OCT system that could perform both automated pupillometry and retinal OCT imaging would thus have great clinical utility.

研究设计

研究类型
Observational
观察模型
Case Only
时间视角
Cross Sectional

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • 未提供

排除标准

  • 未提供

结局指标

主要结局

Diagnostic accuracy of the binocular OCT

时间窗: 6 months

The primary objective will be to determine the diagnostic accuracy of a prototype binocular OCT system for the detection of relative afferent pupillary defect (RAPD).

次要结局

  • Test-retest variability of the binocular OCT pupillometry assessment(6 months)