A comparison of cerebral oxygenation as measured by the O3, the NIRO 200 and the INVOS 5100 Near-Infrared Spectrophotometers
- Conditions
- healthy volunteers
- Registration Number
- JPRN-UMIN000029059
- Lead Sponsor
- Yamagata University
- Brief Summary
The differences in baseline mean rScO2 of MASIMO O3 compared to NIRO 300 and INVOS 5100 were significantly lower (p < 0.05). The estimates for both absolute-agreementand consistency tests for NIRO 300 and INVOS 5100 pairs were 0.74. The estimates for NIRO 300 and MASIMO O3 pair for absolute agreement was 0.43. Estimates for INVOS 5100 and MASIMO O3 were similar to those for NIRO 300 and MASIMO O3 (absolute agreement = 0.48).
- Detailed Description
Not available
Recruitment & Eligibility
- Status
- Complete: follow-up complete
- Sex
- All
- Target Recruitment
- 9
Not provided
disease of respiratory, heart or central nervous system
Study & Design
- Study Type
- Interventional
- Study Design
- Not specified
- Primary Outcome Measures
Name Time Method Comparing with values of regional oxygen saturation(rSO2) measured by three devices
- Secondary Outcome Measures
Name Time Method