A comparison of cerebral oxygenation as measured by the O3, the NIRO 200 and the INVOS 5100 Near-Infrared Spectrophotometers

Not Applicable
Conditions
healthy volunteers
Registration Number
JPRN-UMIN000029059
Lead Sponsor
Yamagata University
Brief Summary

The differences in baseline mean rScO2 of MASIMO O3 compared to NIRO 300 and INVOS 5100 were significantly lower (p < 0.05). The estimates for both absolute-agreementand consistency tests for NIRO 300 and INVOS 5100 pairs were 0.74. The estimates for NIRO 300 and MASIMO O3 pair for absolute agreement was 0.43. Estimates for INVOS 5100 and MASIMO O3 were similar to those for NIRO 300 and MASIMO O3 (absolute agreement = 0.48).

Detailed Description

Not available

Recruitment & Eligibility

Status
Complete: follow-up complete
Sex
All
Target Recruitment
9
Inclusion Criteria

Not provided

Exclusion Criteria

disease of respiratory, heart or central nervous system

Study & Design

Study Type
Interventional
Study Design
Not specified
Primary Outcome Measures
NameTimeMethod
Comparing with values of regional oxygen saturation(rSO2) measured by three devices
Secondary Outcome Measures
NameTimeMethod