Multicenter Diagnostic Clinical Performance Study For Automated Detection of Diabetic Retinopathy
试验速览
- 阶段
- 不适用
- 状态
- 尚未招募
- 发起方
- 入组人数
- 900
- 主要终点
- Sensitivity and Specificity of EyeCheckup to detect more than mild diabetic retinopathy
研究概览
简要总结
This study is a retrospective study to assess the efficacy and accuracy of the EyeCheckup software in screening for diabetic retinopathy in healthcare settings.
详细描述
This is a clinical diagnostic performance study for EyeCheckup with data obtained consecutively at multiple geographically different sites within the US population. These sites include all healthcare settings with a trained ophthalmic camera operator using non-mydriatic cameras that are compatible with EyeCheckup. Moreover, primary endpoints are devised that will be used to validate EyeCheckup to diagnose DR in the primary care and eye care settings. During this color fundus images representing either 4 wide field or 7 standard fields of view will be graded by a qualified professional for diabetic retinopathy according to ETDRS on DRSS scale. The human grading output will be compared with diagnosis generated for diabetic retinopathy by AI software. A comparison of both these grades will be used to determine sensitivity and specificity for the EyeCheckup software for screening for diabetic retinopathy.
研究设计
- 研究类型
- Observational
- 观察模型
- Case Only
- 时间视角
- Retrospective
入排标准
- 年龄范围
- 22 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •A diagnosis of either diabetes type 1 or diabetes type 2
- •22 years of age or older
- •Patients Fundus images must include at least 1 macula centered and 1 optic disc nerve centered image
- •All the subjects must be sequentially photographed meaning no omissions of data between a certain time frame by the sites providing images to ensure consecutive selection
- •No history of any other retinal vascular disease, glaucoma, or another disease that may affect the appearance of the retina or optic disc (refractive error and ocular surface disease are allowed)
- •Subject with image taken by color fundus photography that meets the following requirements:
- •The resolution of image is 1024×1024 pixels or higher
- •The ability to provide a 7 standard or 4 wide field of view
- •Must obtained from either Topcon NW400 or Cannon Model CR-2 AF or CR-2 Plus AF or Optomed Aurora Cameras along with the following data: their age, ethnicity and gender, patient ocular history as evaluated of verified by the clinical sites where data is obtained
排除标准
- •No diagnosis of either diabetes type 1 or diabetes type 2
- •The macula, optic nerve, or other part in the image of color fundus photography is unclear to determine the disease condition as judged by the certified ophthalmologist who is grading the images.
结局指标
主要结局
Sensitivity and Specificity of EyeCheckup to detect more than mild diabetic retinopathy
时间窗: 1 visit (1 day)
Sensitivity and Specificity of EyeCheckup to detect vision-threatening diabetic retinopathy
时间窗: 1 visit (1 day)
次要结局
- Negative Predictive Value(1 visit (1 day))
- Positive Predictive Value(1 visit (1 day))