临床试验/NCT02649842
NCT02649842已完成不适用

Post-Approval Study of the TECNIS® Toric IOL Extended Cylinder Range (ECR), Models ZCT450, ZCT525 and ZCT600

Abbott Medical Optics0 个研究点目标入组 101 人开始时间: 2016年3月21日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
入组人数
101
主要终点
Rate of Severe Visual Distortions

研究概览

简要总结

This study is a prospective, multicenter, bilateral, non-randomized, open-label, clinical study conducted at up to 20 sites in the USA. Subjects were to be bilaterally implanted with one eye having an extended cylinder (high cylinder) toric IOL (model ZCT450, ZCT525 or ZCT600) and the fellow eye having a high cylinder toric IOL (model ZCT450, ZCT525 or ZCT600) or a lower cylinder toric IOL (ZCT300 or ZCT400). This study evaluates the outcomes for subjects implanted with at least one higher cylinder toric IOL.

详细描述

The purpose of this study is to evaluate the performance of the higher cylinder (ZCT450,ZCT525 and ZCT600) toric IOLs. Some subjects had the higher cylinder toric IOLs in both eyes with other subjects having a higher cylinder toric IOL in one eye and the lower cylinder toric IOL in the other eye.

Per the statistical plan, data for the ZCT450/525/600 toric IOLs were to be pooled together and reported as one group.

The primary endpoint is the rate of severe visual distortions reported by subject at 6 months via questionnaire. The analysis population for this endpoint will be bilaterally implanted subjects with a high cylinder IOL (ZCT450/525/600) in the primary eye and a high cylinder IOL (ZCT450/525/600) or a low cylinder IOL (ZCT300/400) in the fellow eye.

There were no formal secondary endpoints. Data from a few other endpoints (rate of IOL repositioning due to IOL misalignment and percent change in cylinder) have been reported. For these endpoints the analysis population contains all subjects implanted with a ZCT450/525/600 in at least one eye.

Adverse events were reported by subject for all subjects with a high cylinder IOL (ZCT450/525/600) in at least one eye.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Other
盲法
None

入排标准

年龄范围
22 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者
是

入选标准

  • •Minimum 22 years of age
  • •Bilateral cataracts
  • •Preoperative keratometric cylinder of at least 2.00 D in both eyes with at least one eye having approximately 3.00 D to 4.75 D of corneal astigmatism.
  • •Ability to understand, read and write English in order to consent to study participation and complete a study questionnaire
  • •Signed informed consent and HIPAA authorization

排除标准

  • •Irregular corneal astigmatism
  • •Any corneal pathology/abnormality other than regular corneal astigmatism
  • •Previous corneal surgery or recent ocular trauma or intraocular surgery that is not resolved/stable or may affect visual outcomes
  • •Any pupil abnormalities
  • •Subjects with conditions associated with increased risk of zonular rupture, including capsular or zonular abnormalities that may lead to IOL decentration
  • •Known ocular disease or pathology that may affect visual acuity or that may require surgical intervention during the course of the study
  • •Concurrent participation or participation during 30 days prior to preoperative visit in any other clinical study
  • •Planned monovision correction
  • •Patient is pregnant, plans to become pregnant or is lactating

研究组 & 干预措施

Extended Cylinder IOL

Experimental

Approved toric intraocular lenses, Model ZCT450, ZCT525 or ZCT600

干预措施: TECNIS Toric Models ZCT450, ZCT525 or ZCT600 (Device)

结局指标

主要结局

Rate of Severe Visual Distortions

时间窗: 6 months

Rate of severe visual distortions based on data from a self administered subject questionnaire

次要结局

未报告次要终点

研究者

申办方类型
Industry
责任方
Sponsor