NCT01672073已完成不适用

An Open, Non-randomised, Study to Provide Data on the Arterial Pressure Wave Form Obtained Through the Proxima 3 in Patients Who Require the Insertion of an Arterial Line

Sphere Medical Ltd.1 个研究点 分布在 1 个国家目标入组 10 人开始时间: 2013年1月最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
10
试验地点
1
主要终点
To confirm that siting Proxima 3 in the arterial line does not affect the pressure wave form

研究概览

简要总结

The purpose of the trial is to confirm that the presence of the Proxima 3 disposable in the arterial line doesn't significantly attenuate the arterial pressure wave form. The principal aim of the trial will be:

• To obtain quantitative data allowing investigation of the pressure wave form obtained in a standard arterial line compared to various arterial lines fitted with the Proxima 3.

详细描述

The purpose of this study is to examine the transmission of the arterial pressure wave form through Proxima 3 disposable attached to a configurable parts of the patient's arterial line. The Proxima device has been tested by Intertek as part of the Proxima 2 project and found to comply to BS EN 60601-1 2nd edition "Medical electrical equipment - Part 1: General requirements for basic safety and essential performance". Additional internal tests have also been carried out for clause 51.103, frequency response, from BS EN 60601-2-34(2001).

This study part of the on-going development of the Proxima system which monitors metabolic parameters in critically ill patients using a disposable sensor array. This study builds upon the four clinical trials carried out in the past two years by Sphere Medical in partnership with the NHS (REC references: 10/H1308/53, 10/H0308/113, 11/SW/0166, 12/SW/0175).

This study forms part of the development process for the Proxima family of devices.

The Proxima 3:

  • Is a disposable multi-parameter microanalyser
  • Is connected into the patient's arterial line.
  • Will measure all or some of blood gases, haematocrit and electrolytes and is connected to a small bedside monitor, which displays the results.
  • Microanalyser can perform an unlimited number of analyses over a period of up to seventy two hours.
  • Will be used as part of a closed system, reducing infection risk and preventing blood loss.

研究设计

研究类型
Observational
观察模型
Case Only
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Male and female patients over the age of 18
  • Patients will give full informed consent to participate in the study before inclusion.
  • All patients who have an arterial line inserted as standard care will be considered for inclusion.

排除标准

  • Refusal of consent to participate
  • The patient is considered by the investigator to be unsuitable for the study

结局指标

主要结局

To confirm that siting Proxima 3 in the arterial line does not affect the pressure wave form

时间窗: 10 minutes maximum

The data traces collected from the arterial pressure waveform monitor will undergo quantitative assessment to determine the effect on the pressure wave, if any, by introducing the Proxima 3 device in the system.

次要结局

未报告次要终点

研究者

发起方
Sphere Medical Ltd.
申办方类型
Industry
责任方
Sponsor