Prospective Randomized Evaluation of a New Fabric Attenuation Device in Endovascular Interventional Radiology II (PREFER II)
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 106
- 试验地点
- 2
- 主要终点
- Percentage Change
研究概览
简要总结
The PREFER II trial primary objective of the trial is to show how much radiation attenuation is provided by XPF thyroid collars in absolute and relative terms.
详细描述
The PREFER II trial has three main objectives. The primary objective of the trial is to show how much radiation attenuation is provided by XPF thyroid collars in absolute and relative terms. The secondary objective of the study is to measure and compare the amount of radiation that interventional radiologists are exposed to per case in an interventional suite setting using FDA approved XPF thyroid collars or standard 0.5 mm Pb equivalent thyroid collars.
The third objective is to assess the operator comfort wearing the XPF protection devices. In order to achieve these objectives, measurements will be performed in 150 consecutive interventional procedures requiring C-arm fluoroscopy performed at BCVI. A prospective randomization will be performed, assigning participants to wear their own standard 0.5 mm Pb equivalent thyroid collar or to wear the FDA approved XPF thyroid collar. Radiation exposure will be monitored with 2 radiation detectors (TLDs, one inside and one outside the thyroid collar) which the participant is required to wear. After each procedure the participant will be asked to rate the comfort of wearing the devices on a scale from 1 to 100.
研究设计
- 研究类型
- Interventional
- 分配方式
- Randomized
- 干预模型
- Parallel
- 主要目的
- Health Services Research
- 盲法
- Single (Outcomes Assessor)
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Interventional Radiologists practicing in Baptist Cardiac & Vascular Institute
- •consecutive endovascular procedure requiring C-arm fluoroscopy
排除标准
- •Interventional Neuroradiologists and Interventional Cardiologists will not be included in this study.
结局指标
主要结局
Percentage Change
时间窗: 1 day
Radiation attenuation. The null hypothesis is that the XPF thyroid collar is equal to the standard 0.5mm Pb lead equivalent thyroid collar with regard to radiation attenuation/transmission in relative (%) terms. The alternative hypothesis is that the XPF thyroid collar is superior or inferior to the standard thyroid collar. Based on our results obtained during the 60 procedures (PREFER trial), 150 procedures are required to demonstrate a significant difference with an alpha level of 5% and a power of 80%. Radiation attenuation percentage is measure by interrogation of the lead attenuates X-rays counter required to be used in all interventional radiology clinical procedures by each physician operator. The interrogation is done by measuring the energy imparted per unit mass in the counter wore by ionizing radiation to matter at the specified point. The International System of Units (SI) unit of absorbed dose is the joule per kilogram.
次要结局
- Wearing Comfort of the Collar on a Scale From 0-100.(1 day)
- Radiation Exposure (uSv) for for Both the Thyroid and Equivalent Collars(1 day)
- Radiation Exposure (uSv) for XPF Caps and the Equivalent Caps(1 day)
研究者
Raul Herrera, MD
Director, Division of Clinical Research & Outcomes
Baptist Health South Florida