跳至主要内容
临床试验/NCT05100797
NCT05100797Unknown不适用

Validity of the Vyntus CPX Metabolic Cart for Assessing Postprandial Energy Expenditure and Respiratory Exchange Ratio

Universidad de Granada1 个研究点 分布在 1 个国家目标入组 12 人开始时间: 2021年12月20日最近更新:
适应症

试验速览

阶段
不适用
入组人数
12
试验地点
1
主要终点
VO2 post-prandial gas exchange assessed by indirect calorimetry

研究概览

简要总结

The aims of the present study are described below:

Aim 1: In-vitro measures to determine the validity, precision and accuracy of the Vyntus CPX metabolic cart assessed by:

1.1. Methanol burning tests. 1.2. Pure gas (nitrogen [N2] and carbon dioxide [CO2]) infusions to simulate both, constant and variable fluxes of oxygen consumption (VO2) and carbon dioxide production (VCO2) in different conditions and moments.

  1. Immediately after or prior the methanol burning test (in a randomized order).
  2. In separate days and time-periods.

Aim 2: In-vivo measures using the Vyntus CPX metabolic cart:

2.1. To assess the resting metabolic rate (RMR or resting energy expenditure [REE]) and respiratory exchange ratio (RER) day-to-day biological reproducibility (i.e. reliability) of the metabolic cart.

2.2. Further, the investigators will use an oral glucose tolerance test (OGTT) to determine the post-prandial day-to-day biological reproducibility (especifically the post-prandial RER) of the metabolic cart.

详细描述

The study will be developed in three different experiments which are detailed below:

EXPERIMENT 1 (AIM 1.1 AND 1.2.A):

The investigators will determine the Vyntus CPX measurement error over 14 different days by using both, methanol burnings and pure gas infusions. The order will be random.

  1. Methanol burning: the investigators will burn the methanol inside the glass-cage and the investigators will measure dynamically the weight of burned alcohol; gas exchange will be continuously measured with the Vyntus CPX while the produced gases are directed into the metabolic cart's hose tube. The burning will take approximately 25-30 minutes and is expected to produce a RER=0.667.
  2. Pure gas infusions: Pure N2 (concentration ≥99.9997%) for diluting ambient O2 and CO2 (concentration ≥99.995%) will be infused using high-precision mass-flow controllers (358 Series, Analyt-MTC, Müllheim, Germany; 0-2 l/min) into the metabolic cart's hose tube. The infused volumes of N2 and CO2 will be selected to replicate (i.e. mimic) as much as possible the VO2 and VCO2 produced by the methanol burning test. The gas infusions will last 20 minutes.

EXPERIMENT 2 (AIM 1.2.B):

研究设计

研究类型
Observational
观察模型
Other
时间视角
Other

入排标准

年龄范围
18 Years 至 40 Years(Adult)
性别
All
接受健康志愿者

入选标准

  • Age 18-40 y.
  • BMI between 18.5 kg/m2 and 30 kg/m2 (both inclusive).
  • Willing to fast and not to consume non-caloric stimulant for 12h before the assessments.
  • Willing to avoid moderate or vigorous physical activity during 24-48h respectively.
  • Have a stable body weight over the last 3 months (changes ≤3 kg) and not being enrolled in a weight loss program.

排除标准

  • Smoking or use of tobacco products within the last 3 months.
  • Diagnosed with diabetes.
  • Previous bariatric surgery (or other surgeries) for obesity or weight loss.
  • Use of medications affecting metabolism or sleep.
  • History of neurological disease.
  • History of cardiovascular disease, including hypertension, or other chronic diseases.
  • Pregnant, planning to become pregnant, or breastfeeding.
  • Adherence to special restrained diets (e.g., low-carbohydrates, low-fat, or vegetarian/vegan diets) over the last 3 months.
  • Claustrophobia impeding to stay calm under a plastic canopy.

结局指标

主要结局

VO2 post-prandial gas exchange assessed by indirect calorimetry

时间窗: 5 hours

Volume of oxygen consumption (VO2 \[ml/min\])

VCO2 post-prandial gas exchange assessed by indirect calorimetry

时间窗: 5 hours

Volume of carbon dioxide production (VCO2 \[ml/min\])

VO2 validity of the metabolic cart. In-vitro measurements and/or simulations to determine the validity, precision and accuracy of the metabolic cart assessed by methanol burning tests and controlled pure gas infusions.

时间窗: 14 separate days; measurements of 2 hours per day

Volume of oxygen consumption (VO2) measurement error (ml/min)

CHOox post-prandial gas exchange assessed by indirect calorimetry

时间窗: 5 hours

Carbohydrate oxidation (CHOox \[g/min\])

EE post-prandial gas exchange assessed by indirect calorimetry

时间窗: 5 hours

-Energy Expenditure (EE \[kcal/day\])

VCO2 validity of the metabolic cart. In-vitro measurements and/or simulations to determine the validity, precision and accuracy of the metabolic cart assessed by methanol burning tests and controlled pure gas infusions.

时间窗: 14 separate days; measurements of 2 hours per day

Volume of carbon dioxide production (VCO2) measurement error (ml/min)

EE validity of the metabolic cart. In-vitro measurements and/or simulations to determine the validity, precision and accuracy of the metabolic cart assessed by methanol burning tests and controlled pure gas infusions.

时间窗: 14 separate days; measurements of 2 hours per day

Energy Expenditure (EE) measurement error (kcal/day)

RER post-prandial gas exchange assessed by indirect calorimetry

时间窗: 5 hours

Respiratory Exchange Ratio (RER \[calculated as VCO2-to-VO2 ratio\])

FATox post-prandial gas exchange assessed by indirect calorimetry

时间窗: 5 hours

Fat oxidation (FATox \[g/min\])

RER validity of the metabolic cart. In-vitro measurements and/or simulations to determine the validity, precision and accuracy of the metabolic cart assessed by methanol burning tests and controlled pure gas infusions.

时间窗: 14 separate days; measurements of 2 hours per day

Respiratory Exchange Ratio (RER) measurement error

次要结局

  • Anthropometric measurements 1, weight(1 day)
  • Anthropometric measurements 2, height(1 day)
  • Anthropometric measurements 3, body mass index(1 day)
  • Post-prandial blood glucose concentrations(5 hours)
  • Body composition outcomes (DXA scann)(1 day)

研究者

申办方类型
Other
责任方
Principal Investigator
主要研究者

Jonatan Ruiz Ruiz

Principal Investigator

Universidad de Granada

研究点 (1)

Loading locations...