简要总结
The purpose of this trial is to determine the accuracy of the Blood Pressure (BP) measurement of multifunction KEITO devices under test requirements of the Standard American National Standards Institute(ANSI)/Association for the Advancement of Medical Instrumentation(AAMI)/International Organization for Standardization (ISO): ANSI/AAMI/ISO 81060-2:2009.
详细描述
Nowadays is important controlling our basic health parameters. Due this, the multifunction KEITO Devices was designed to offer to the general public a convenient way to do so easily in the public sites.
One of the most important parameter is the Blood Pressure (BP)measurement. To warrant their efficacity and accuracy, was essential to perform this trial to ensure that the measurements obtained meet with the requirements of the specific U.S. standard about Non-invasive sphygmomanometers.
By Validating with auscultatory reference sphygmomanometer, as defined on the ANSI/AAMI/ISO 81060-2:2009 , we have been intending to demonstrate that the multifunction KEITO devices, particularly about the Blood Pressure (BP) Measurement, have the same efficacy and accuracy as the conventional auscultatory sphygmomanometers.
To include most of the groups of Blood Pressure (BP) measurements types on this clinical trial, a great variety of BP ranges have been selected, always in accordance with the standard.
研究设计
- 研究类型
- Interventional
- 分配方式
- Non Randomized
- 干预模型
- Crossover
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 83 Years(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Wrist circumference: 12 to 20 Cm
- •At least 5% of the readings shall have Systolic BP<=100 mmHg
- •At least 5% of the readings shall have Systolic BP>=160 mmHg
- •At least 20% of the readings shall have Systolic BP>=140 mmHg
- •At least 5% of the readings shall have Diastolic BP<=60 mmHg
- •At least 5% of the readings shall have Diastolic BP>=100 mmHg
- •At least 20% of the readings shall have Diastolic BP>=85 mmHg
排除标准
- •Pregnant Women
结局指标
主要结局
BP Measurements Using the Reference Auscultatory Sphygmomanometer
时间窗: Five minutes
The subject was measured by the two observers at the same time using the reference sphygmomanometer and double stethoscope. The observers agree to take the K5 korotkoff sound. Each observer takes notes of their own measures.
BP Measurement Using Multifunction KEITO
时间窗: 30 seconds
Prior starting the measurement phase using Multifunction KEITO, the under test device was reset before each new measurement. The subject has to be at rest at least 60 seconds before be measured.
次要结局
未报告次要终点