Observation of ImageReady™ MR Conditional Defibrillation System in China (MR ICD)

Boston Scientific Corporation1 个研究点 分布在 1 个国家目标入组 20 人开始时间: 2018年4月3日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
20
试验地点
1
主要终点
The Percentage of Participant Whose the Average RV Shocking Impedance is >200 Ohm 1 Month Post Scan, While it is ≤ 200 Ohm Before Scan

研究概览

简要总结

To observe the safety and effectiveness of ImageReady™ MR conditional defibrillation system in a Chinese population.

详细描述

This is a multi-center, prospective, single-arm study, aimed to observe the safety and effectiveness of ImageReady™ MR conditional defibrillation system in a Chinese population.It will enroll 20 subjects in 5 sites in China.The primary safety endpoint is MR scan-related Complication-free rate between the MR Scan and the MRI + 1 Month Visit. The primary effectiveness endpoints include:1.Abnormal increase in RV shocking impedance from the pre-MR scan to the 1 Month post-MR scan. 2.Increase in RV pacing threshold from the pre-MR scan to the 1 Month post-MR scan. 3.Decrease in RV sensed amplitude from the pre-MR scan to the 1 Month post-MR scan. 4.Increase in LV pacing threshold from the pre-MR scan to the 1 Month post-MR scan. 5.Decrease in LV sensed amplitude from the pre-MR scan to the 1 Month post- MR scan.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Treatment
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Subject is indicated to Class I/II indications per guidelines/consensus released by Chinese Society of Cardiac Pacing and Electrophysiology
  • Subject must have the ImageReady System as their initial (de novo) defibrillation system implant
  • Subject will receive an ICD or CRT-D pulse generator in the left or right pectoral region ICD
  • Subject is able and willing to undergo an MR scan
  • Subject is willing and capable of providing informed consent and participating in all testing/ visits associated with this clinical study at an approved clinical study center and at the intervals defined by this protocol
  • Subject is age 18 or above

排除标准

  • Subject has other active or abandoned implanted cardiac rhythm devices, components or accessories present such as pulse generators, leads, lead adaptors or extenders
  • Presence of metallic objects that represent a contraindication to MR imaging at the discretion of the Radiologist and impacting the ability to conduct the study protocol
  • Subject needs or will need a medically necessary MR scan, before completing the 1-month post-MR follow-up visit
  • Subject with:
  • A history of syncope related to brady-arrhythmia
  • A history of syncope of unknown etiology
  • Sinus pauses (Pause > 2 s)
  • Permanent or intermittent complete AV block
  • Documentation of progressive AV nodal block over time
  • Trifascicular block (alternating bundle branch block or PR > 200 ms with LBBB or other bifascicular block)
  • Note: It is required to run a 12 lead ECG and a 10s rhythm strip to document this exclusion criterion.
  • Subject is not clinically capable of tolerating the absence of pacing or Resynchronization therapy support in a supine position for the duration that the pulse generator is in MRI Protection Mode, per Physician discretion
  • Subject is not clinically capable of tolerating the absence of Tachycardia therapy support for the duration that the pulse generator is in MRI Protection Mode, per Physician discretion
  • Subjects currently requiring dialysis
  • Subject has a mechanical heart valve
  • Subject has a known or suspected sensitivity to dexamethasone acetate (DXA)
  • Subject is currently on the active heart transplant list
  • Subject has documented life expectancy of less than 12 months
  • Subject is enrolled in any other concurrent study that might interfere with this study
  • Women of childbearing potential who are or might be pregnant at the time of this study

结局指标

主要结局

The Percentage of Participant Whose the Average RV Shocking Impedance is >200 Ohm 1 Month Post Scan, While it is ≤ 200 Ohm Before Scan

时间窗: MRI+1 Month visit ( 10~13 weeks from 0 day)

The normal RV shocking impedance measured by the system should be≤200 Ohm, with \>200 Ohm considered abnormal. The primary effectiveness endpoint 1 is defined as that the average RV shocking impedance is \>200 Ohm at 1 month post scan, while it is ≤ 200 Ohm before scan.There is no hypothesis testing in the MR ICD study, and it can be considered reasonable if up to 2 failed cases of primary effectiveness endpoint 1 occur.

The Percentage of Participant That Have the Average RV Sensed Amplitude at the MRI + 1 Month Visit Remains < 5.0 mV or Less Than 50% of the Pre-MR Scan Value

时间窗: MR+1 Month visit( 10~14 weeks from 0 day)

Subjects will be considered a success if the average sensed amplitude at the MRI + 1 Month Visit remains ≥ 5.0 mV and above 50% of the pre-MR scan value, otherwise, it is a primary effectiveness endpoint 3 event.The performance goal of this endpoint for the ENABLE MRI study was 85%. There is no hypothesis testing in the MR ICD study, and it can be considered reasonable if up to 2 failed cases of primary effectiveness endpoint 3 occur.

The Percent of Participants Who Were MR Scan-Related ImageReady System Complication -Free

时间窗: MRI + 1 Month Visit(10-13 weeks from 0 day)

The primary safety endpoint of the MR ICD study will be assessed for all subjects who undergo any portion of the study-required MR scan sequences. Safety will be confirmed by evaluating the MR scan related ImageReady System complication-free rate (CFR) between the MR Scan and the MRI Visit + 1 Month.MR scan related ImageReady System complication is relation with ImageReady System and MR Scan.

The Percentage of Participant That Have an Increase in Average RV Pacing Thresholds > 0.5V (at 0.5 ms) From Pre-MR Scan to MRI Visit + 1 Month Follow-up

时间窗: MRI+1Month visit( 10~14 weeks from 0 day)

Subjects that have an increase in average RV pacing thresholds ≤ 0.5V (at 0.5 ms) from pre-MR Scan to MRI Visit + 1 Month follow-up will be considered a success, otherwise, it is a primary effectiveness endpoint 2 event.The performance goal of this endpoint for the ENABLE MRI study was 87%. There is no hypothesis testing in the MR ICD study, and it can be considered reasonable if up to 2 failed cases of primary effectiveness endpoint 2 occur.

The Percentage of Participant Have an Increase in Average LV Pacing Thresholds >1.0V (at 0.5 ms) From Pre-MR Scan to MRI Visit + 1 Month Follow-up

时间窗: MR +1 Month visit( 10~14 weeks from 0 day)

Subjects that have an increase in average LV pacing thresholds ≤1.0V (at 0.5 ms) from pre-MR Scan to MRI Visit + 1 Month follow-up will be considered a success, otherwise, it is a primary effectiveness endpoint 4 event.The performance goal of this endpoint for the ENABLE MRI study was 87%. There is no hypothesis testing in the MR ICD study, and it can be considered reasonable if up to 2 failed cases of primary effectiveness endpoint 4 occur.

the Percentage of Participant That Have the Average LV Sensed Amplitude at the MRI + 1 Month Visit Remains <5.0 mV or Above 50% of the Pre-MR Scan Value

时间窗: MR +1 Month visit( 10~14 weeks from 0 day)

Subjects will be considered a success if the average sensed amplitude at the MRI + 1 Month Visit remains ≥ 5.0 mV and above 50% of the pre-MR scan value, otherwise, it is a primary effectiveness endpoint 5 event.The performance goal of this endpoint for the ENABLE MRI study was 85%. There is no hypothesis testing in the MR ICD study, and it can be considered reasonable if up to 2 failed cases of primary effectiveness endpoint 5 occur.

次要结局

未报告次要终点

研究者

申办方类型
Industry
责任方
Sponsor