Evaluation of the Tof Cuff for Perioperative Neuromuscular Transmission Monitoring During Recovery of Moderate and Deep Neuromuscular Block Compared to Acceleromyography and Electromyography

Leiden University Medical Center2 个研究点 分布在 1 个国家目标入组 250 人开始时间: 2017年5月1日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
250
试验地点
2
主要终点
Depth of Neuromuscular Block During Moderate Neuromuscular Blockade (Outcome Measure/Row Title Train of Four Ratio) and During Deep Neuromuscular Blockade (Outcome Measure/ Row Title Post Tetanic Count)

研究概览

简要总结

Acceleromyography (AMG) is the most wide spread used method to assess neuromuscular block during anesthesia. However AMG is known to be inaccurate when compared to the gold standard in neuromuscular transmission monitoring, electromyography (EMG). Furthermore when the patients arms require to be positioned next to the body and beneath surgical drapes, AMG measurements are often hindered and inaccurate. The TOF cuff is a new device which measures neuromuscular blockade at the upper arm with a blood pressure cuff. It overcomes the previously mentioned disadvantages of AMG. However, it validity compared to EMG and AMG has not yet fully been investigated.

This study aims to compare the bias, limits of agreement and precision of the Train-of-Four cuff relative to AMG and EMG during recovery of moderate and deep neuromuscular block in patients with normal body mass index and morbidly obese patients.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • American society of Anesthesiologist Physical Status class I-III
  • >18 years of age
  • Ability to give oral and written informed consent

排除标准

  • Known or suspected neuromuscular disorders impairing neuromuscular function;
  • Allergies to muscle relaxants, anesthetics or narcotics;
  • A (family) history of malignant hyperthermia;
  • Women who are or may be pregnant or are currently breast feeding;
  • Renal insufficiency, as defined by a glomerular filtration rate < 30 ml/min
  • Scheduled for anesthesia without the use of muscle relaxants.

结局指标

主要结局

Depth of Neuromuscular Block During Moderate Neuromuscular Blockade (Outcome Measure/Row Title Train of Four Ratio) and During Deep Neuromuscular Blockade (Outcome Measure/ Row Title Post Tetanic Count)

时间窗: at 5 minute intervals during the length of the entire procedure [which lasted between 20 to 372 minutes]; a mean of the differences between the TOF-Cuff and electromyography of all measurements is calculated and reported below

Depth of as neuromuscular block will be compared between the Tof-Cuff and electromography at five minute intervals during moderate neuromuscular blockade. Bland-Altman analysis modified for repeated measurements (http://sec.lumc.nl/method_agreement_analysis). This Bland-Altman analysis corrects for between subject variability of repeated paired measurements in individual subjects. Bland-Altman analysis estimates bias and limits of agreement (95% differences between compared devices) between Tof-Cuff and electromography and evaluates instrumental imprecision by calculating the repeatability coefficient, which is equal to the standard deviation of the within-subject variability of each device. For interpretation of the Bland-Altman bias during offset of neuromuscular blockade one can assume that a bias above zero indicates that TOF-Cuff overestimates neuromuscular blockade recovery whilst a bias below zero indicates that Tof-Cuff underestimates neuromuscular blockade recovery.