跳至主要内容
临床试验/NCT02041702
NCT02041702已完成不适用

A Clinical Evaluation of the Safety and Efficacy of St. Jude Medical (SJM) Magnetic Resonance Imaging (MRI) Conditional Pacing System in Patients Undergoing Cardiac MRI

Abbott Medical Devices13 个研究点 分布在 5 个国家目标入组 283 人开始时间: 2014年6月9日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
入组人数
283
试验地点
13
主要终点
Freedom From MRI Scan-related Complications

研究概览

简要总结

The aim of this study was to assess safety and efficacy of SJM MRI conditional pacing system. The study hypothesized that cardiac MRI scan would not affect performance of SJM MRI conditional pacing system.

详细描述

Accent Cardiac MRI study was a prospective, randomized ,multi-center study conducted in Asia. Subjects indicated for dual-chamber pacemaker and who signed study consent form were considered enrolled and post implant were randomized to either Cardiac MRI Scan Group or Control Group.

Clinical evaluation and device interrogation were performed at baseline, pre- and post-MRI scan, and one month post-MRI for all subjects. At 9-12 weeks post-implant, patients in Cardiac MRI Scan Group underwent an elective non-diagnostic cardiac MRI scan at 1.5 Tesla (T) while patients in Control group underwent device interrogation and clinical evaluation twice with a 45-minute waiting period in between. The safety endpoint was freedom from MRI-related complications while efficacy endpoint was change in right atrial/ ventricular (RA/RV) capture threshold and sensing amplitude from pre-MRI scan to one month post-MRI scan.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Parallel
主要目的
Supportive Care
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • 未提供

排除标准

  • 未提供

研究组 & 干预措施

Control Group

No Intervention

Cardiac MRI Scan Group

Experimental

干预措施: Non-Diagnostic Cardiac MRI scan (Diagnostic Test)

结局指标

主要结局

Freedom From MRI Scan-related Complications

时间窗: MRI Visit ,1 Month Post MRI Visit

Number of subjects who were free from MRI scan-related complications

Change in Right Atrial (RA) Capture Threshold @0.5 Millisecond (ms) From Pre-MRI Scan to 1 Month After MRI Scan

时间窗: MRI Visit ,1 Month Post MRI Visit

Number of subjects who experienced an increase in RA capture threshold @ 0.5 ms at 1-Month post MRI scan visit ≤ 0.5 Volt (V) compared to pre-MRI scan value collected at MRI scan visit

Change in Right Ventricular (RV) Capture Threshold @0.5ms From Pre-MRI Scan to 1 Month After MRI Scan

时间窗: MRI Visit, 1 Month Post MRI Visit

Number of subjects who experienced an increase in RV capture threshold @ 0.5 ms at 1-Month post MRI scan visit ≤ 0.5 V compared to pre-MRI scan value collected at MRI scan visit

Change in RA Sensing Amplitude From Pre-MRI Scan to 1 Month After MRI Scan

时间窗: MRI Visit, 1 Month Post MRI Visit

Number of subjects who experienced a decrease in RA sensing amplitude ≤ 50% and at least 1.5 mV at 1-Month post MRI scan visit compared to pre-MRI scan value collected at MRI scan visit

Change in RV Sensing Amplitude From Pre-MRI Scan to 1 Month After MRI Scan

时间窗: MRI Visit, 1 Month Post MRI Visit

Number of subjects who experienced a decrease in RV sensing amplitude ≤ 50% and at least 5 mV at 1-Month post MRI scan visit compared to pre-MRI scan value collected at MRI scan visit

次要结局

未报告次要终点

研究者

申办方类型
Industry
责任方
Sponsor

研究点 (13)

Loading locations...

相似试验