临床试验/NCT03149042
NCT03149042已完成不适用

Validation of a Computed Tomography (CT) Based Fractional Flow Reserve (FFR) Software Using the 320 Detector Aquilion ONE CT Scanner.

State University of New York at Buffalo1 个研究点 分布在 1 个国家目标入组 75 人开始时间: 2016年5月28日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
75
试验地点
1
主要终点
Comparison of CT Based FFR With Invasive FFR, Correlation Analysis

研究概览

简要总结

Coronary Computed Tomography Angiography (CCTA) contrast opacification gradients and FFR-CT estimation can aid in the severity estimation of significant atherosclerotic lesions. Currently, FFR-CT algorithms can only be optimized using theoretical models and can only be validated in large multi-center clinical trials. Using patient specific 3D printed coronary phantoms would allow optimization of FFR-CT algorithms with a measured validation technique without the need for large clinical trials. Thus the investigators believe that this study will result in a FFR-CT algorithm/method with a better predictability for arterial lesion severity than those existing on the market today. Flow measurements will be compared with: CT-FFR for both patients and phantoms, angio lab FFR measurements and 30 days follow-up. This pilot clinical study includes ~50 patients over a year and half at GVI.

详细描述

Coronary Computed Tomography Angiography (CCTA) contrast opacification gradients and FFR-CT estimation can aid in the severity estimation of significant atherosclerotic lesions. Following this trend, the investigators recently developed a collaboration between Brigham and Women's Hospital (BWH) and Gates Vascular Institute (GVI). The investigators 3D-printed patient specific coronary phantoms at (GVI) and scanned them with a Toshiba Aquilion scanner to test several aspects of the contrast opacification gradients using a method established at BWH. The initial results showed strong correlation between the flow in the phantom and opacification gradients. The investigators believe that this approach could be further developed to test and validate FFR-CT algorithms. Currently, FFR-CT algorithms can only be optimized using theoretical models and can only be validated in large multi-center clinical trials. This phantom approach would allow optimization of FFR-CT algorithms with a measured validation technique without the need for large clinical trials. Thus the investigators believe that this study will result in a FFR-CT algorithm/method with a better predictability for arterial lesion severity than those existing on the market today. The approach is to use the infrastructure at GVI to perform a detailed validation of the FFR-CT method using 3D printed patient specific phantoms. The subject enrollment criteria is: at least one CCTA, at least one lesion with >50% stenosis or 30-50% and an angio based FFR. Each patient will have a 3D phantom printed, containing the culprit lesion and used in a benchtop flow analysis. Flow measurements will be compared with: CT-FFR for both patients and phantoms, angio lab FFR measurements and 30 days follow-up. This pilot clinical study will include ~50 patients over a year and half at GVI. The investigators are confident that this approach performed via 3D-phantom testing will prove the validity of FFR-CT based measurements as well as develop a new standard for validating FFR-CT algorithms.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • All the patients >18 yrs of age , who are undergoing CCTA and angio-FFR. Patients who are (1) scheduled for clinically mandated elective invasive coronary angiography (ICA) at Buffalo General Hospital or (2) clinically mandated CTA will be screened.

排除标准

  • Adults unable to consent
  • Individuals who are not yet adults (infants, children, teenagers)
  • Pregnant women
  • Prisoners
  • atrial fibrillation,
  • Renal insufficiency (estimated glomerular filtration rate (GFR) <60 ml/min/1.73 m2),
  • Active Bronchospasm prohibiting the use of beta blockers
  • Morbid obesity (body mass index 40 kg/m2)
  • Contraindications to iodinated contrast.
  • Emergencies requiring immediate intervention or patients unable to consent.
  • Patients not showing coronary calcium during Calcium Scoring procedures

结局指标

主要结局

Comparison of CT Based FFR With Invasive FFR, Correlation Analysis

时间窗: 24 hours

Patient CCTA images were imported into Vitrea segmentation software (Vital Images, Minnetonka, MN) for use in the research-based CT based FFR algorithm. The software analyzes four data volumes acquired a 70%, 80%, 90% and 99% of the R-R interval and computes the FFR based on the changes in vessel diameter and computational fluid dynamics. Within the algorithm, the aortic root and three main coronary arteries (LAD, LCX, and RCA) were automatically segmented, and then manually adjusted to obtain accurate centerline and contours. The CT based FFR was calculated and the user adjusted the location of the distal pressure measurement to calculate the CT basedFFR at the same location as Invasive-FFR, two lesion lengths below the distal end of the lesion. Pearson Correlation between Invasive FFR and CT based FFR was measured

Comparison of CT Based FFR With Invasive FFR, ROC Analysis

时间窗: 24 hours

Patient CCTA images were imported into Vitrea segmentation software (Vital Images, Minnetonka, MN) for use in the research-based CT based FFR algorithm. The software analyzes four data volumes acquired a 70%, 80%, 90% and 99% of the R-R interval and computes the FFR based on the changes in vessel diameter and computational fluid dynamics. Within the algorithm, the aortic root and three main coronary arteries (LAD, LCX, and RCA) were automatically segmented, and then manually adjusted to obtain accurate centerline and contours. The CT based FFR was calculated and the user adjusted the location of the distal pressure measurement to calculate the CT basedFFR at the same location as Invasive-FFR, two lesion lengths below the distal end of the lesion. Area under the Receiver Operator Characteristic were measured where an Invasive FFR\<=0.8 was considered positive.

Comparison of CT Based FFR With Invasive FFR, Sensitivity

时间窗: 24 hours

Patient CCTA images were imported into Vitrea segmentation software (Vital Images, Minnetonka, MN) for use in the research-based CT based FFR algorithm. The software analyzes four data volumes acquired a 70%, 80%, 90% and 99% of the R-R interval and computes the FFR based on the changes in vessel diameter and computational fluid dynamics. Within the algorithm, the aortic root and three main coronary arteries (LAD, LCX, and RCA) were automatically segmented, and then manually adjusted to obtain accurate centerline and contours. The CT based FFR was calculated and the user adjusted the location of the distal pressure measurement to calculate the CT basedFFR at the same location as Invasive-FFR, two lesion lengths below the distal end of the lesion. Sensitivity were measured where an Invasive FFR\<=0.8 was considered positive. Sensitivity reflects the percentage of true positive cases identified by CT-FFR compared to I-FFR

Comparison of CT Based FFR With Invasive FFR, Specificity

时间窗: 24 hours

Patient CCTA images were imported into Vitrea segmentation software (Vital Images, Minnetonka, MN) for use in the research-based CT based FFR algorithm. The software analyzes four data volumes acquired a 70%, 80%, 90% and 99% of the R-R interval and computes the FFR based on the changes in vessel diameter and computational fluid dynamics. Within the algorithm, the aortic root and three main coronary arteries (LAD, LCX, and RCA) were automatically segmented, and then manually adjusted to obtain accurate centerline and contours. The CT based FFR was calculated and the user adjusted the location of the distal pressure measurement to calculate the CT basedFFR at the same location as Invasive-FFR, two lesion lengths below the distal end of the lesion. Specificity was measured, where an Invasive FFR\<=0.8 was considered positive. Specificity reflects the percentage of true negative cases identified by CT-FFR compared to I-FFR

次要结局

  • Comparison of Bench-top FFR Using 3D Printed Patient Specific Phantoms With Invasive FFR, Pearson Correlation(4 weeks from baseline)
  • Comparison of Bench-top FFR Using 3D Printed Patient Specific Phantoms With Invasive FFR, Specificity(4 weeks from baseline)
  • Comparison of CT Based FFR With Bench-top FFR Using 3D Printed Patient Specific Phantoms(4 weeks from baseline)
  • Comparison of Bench-top FFR Using 3D Printed Patient Specific Phantoms With Invasive FFR, ROC Analysis(4 weeks from baseline)
  • Comparison of Bench-top FFR Using 3D Printed Patient Specific Phantoms With Invasive FFR, Sensitivity(4 weeks from baseline)

研究者

申办方类型
Other
责任方
Principal Investigator
主要研究者

Ciprian Ionita

Principal Investigator

State University of New York at Buffalo

研究点 (1)