Safety and Performance Study of Large Hole Vascular Closure Device - FRONTIER II Study
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 65
- 试验地点
- 12
- 主要终点
- Primary Endpoint
研究概览
简要总结
The purpose of this Clinical Investigation is to validate that the clinical use of the VIVASURE CLOSURE DEVICE™ is safe for the operator, patient and third parties, and to confirm its performance to percutaneously close femoral arterial puncture sites in the range of 18-24 F, post endovascular procedures.
This is a non-inferiority study based on safety. Safety will be assessed by incidence and severity of major complication rates directly related to the VIVASURE CLOSURE DEVICE™ up to 3 months from implantation is no worse than those associated with cut-down and sutured close.
详细描述
This study will be a prospective, multi-centred, non-randomized study to investigate the safety and performance of the VIVASURE CLOSURE DEVICE™. The study shall not be blinded prior to, during or post the procedure. All patients undergoing a procedure requiring an arteriotomy in the range of 18 to 24 F, via the common femoral artery will be screened against the inclusion/exclusion criteria. If the patient meets the requirements of the clinical investigation, they shall be invited to participate, provide informed consent and shall subsequently be assigned a subject number.
All subjects shall have an immediately post procedure, 24 hour, 1, 3 and 12 month follow-up assessment. Safety data from the follow-ups will be assessed by the Data Safety Monitoring Committee.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Treatment
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Over 18 years of age
- •Each patient, or his or her guardian or legal representative, is willing to give informed consent
- •Clinically indicated for an endovascular procedure involving access through the femoral artery, with an access puncture of 18 - 24 French (F)
- •Females who are not pregnant or lactating, and not planning to become pregnant in ≤ 12 months
排除标准
- •Severe acute non-cardiac systemic disease or terminal illness with a life expectancy of less than one year
- •Evidence of systemic bacterial or cutaneous infection
- •Evidence of MRSA (Methicillin-resistant Staphylococcus aureus) and/or VRE (vancomycin-resistant Enterococci) colonisation
- •Arterial access other than the common femoral artery
- •Patients suffering with definitive or potential coagulopathy or platelet count less than 100,000/μl
- •Patient with a haematocrit of less than 32 %
- •A measured activated clotting time (ACT) of greater than 350 seconds immediately prior to sheath removal
- •If patients are expected to be continuously treated with anticoagulation therapy post-procedure such that their ACT reading is expected to be elevated above 350 seconds for more than 24 hours after the procedure
- •Evidence of arterial diameter stenosis greater than 20 % within 20 mm of the arteriotomy
- •Circumferential calcification within 20 mm of the arteriotomy
- •Use of systemic thrombolytic agents within 24 hours prior to or during the catheterisation procedure which cause the concentration of fibrinogen to be less than 100 mg/dl
- •Patients in which the arteriotomy is less than 18 F or greater than 24 F
- •Known allergy to any of the materials used in the device
- •Currently enrolled in any other investigational clinical study, whereby the primary endpoint has not yet been achieved
- •Patients judged unsuitable for surgical repair of the access site
- •If puncture site is via a vascular graft
- •If there is any indication that the puncture has been made in the profunda femoris or located less than 10 mm above the profunda femoris
- •Patients with a common femoral artery lumen diameter of less than 7 mm
- •Patients that have a lower extremity amputation from the ipsilateral or contralateral limb
- •Patients that have undergone a percutaneous procedure using a non-absorbable vascular closure device (excluding suture mediated) for haemostasis in the ipsilateral leg
- •Patients that have undergone a percutaneous procedure greater than 8 F in the ipsilateral leg, within the previous 90 days
- •Patients that have undergone a percutaneous procedure of 8 F or less using an absorbable intravascular closure device for haemostasis, in the ipsilateral leg, within the previous 90 days
- •Patients that have undergone a percutaneous procedure of 8 F or less using a suture mediated closure device for haemostasis, in the ipsilateral leg, within the previous 30 days
- •Patients that have undergone a percutaneous procedure of 8 F or less using manual/mechanical pressure for haemostasis in the ipsilateral leg, within the previous 30 days
- •Patients with an acute haematoma of any size, arteriovenous fistula or Pseudoaneurysm at the access site
- •Significant blood loss/transfusion during interventional procedure or within 20 days of procedure requiring transfusion of greater than 4 units of blood
- •Angiographic evidence of arterial laceration, dissection or stenosis within the external iliac or femoral artery before the use of the VCD
- •Severe claudication, stenosis of the iliac artery > 50% or previous bypass surgery/stent placement in the region of the vascular access
研究组 & 干预措施
Treatment
VIVASURE CLOSURE DEVICE™
干预措施: VIVASURE CLOSURE DEVICE™ (Device)
结局指标
主要结局
Primary Endpoint
时间窗: 3 months
Incidence and severity of major complication rates (as defined by VARC-2) directly related to the VIVASURE CLOSURE DEVICE™ up to 3 months from implantation, is no worse than those associated with cut-down and sutured close
次要结局
- Secondary Endpoint(3 Months)